Ariz. Admin. Code § R9-7-714 - Authorization for Calibration, Transmission, and Reference Sources
Any person authorized by R9-7-703 for medical use of radioactive material may receive, possess, and use any of the following radioactive material for check, calibration, transmission, and reference use.
1. Sealed sources, not exceeding
1.11 GBq (30 mCi) each, manufactured and distributed by a person licensed under
Article 3 of this Chapter or equivalent NRC or Agreement State
regulations.
2. Sealed sources, not
exceeding 1.11 GBq (30 mCi) each, redistributed by a licensee authorized to
redistribute the sealed sources manufactured and distributed by a person
licensed under Article 3 of this Chapter, providing the redistributed sealed
sources are in the original packaging and shielding and are accompanied by the
manufacturer's approved instructions.
3. Any radioactive material with a half-life
not longer than 120 days in individual amounts not to exceed 0.56 GBq (15
mCi).
4. Any radioactive material
with a half-life longer than 120 days in individual amounts not to exceed the
smaller of 7.4 MBq (200 µCi) or 1000 times the quantities in Article 4,
Appendix B of this Chapter.
5.
Technetium-99m in amounts as needed.
6. A licensee is limited to five sources of
radiation authorized under subsections (1) through (3), unless otherwise
specified in the licensee's radioactive material license.