Ariz. Admin. Code § R9-7-743 - Therapy-related Computer Systems

The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing shall include, as applicable, verification of:

1. The source-specific input parameters required by the dose calculation algorithm;
2. The accuracy of dose, dwell time, and treatment time calculations at representative points;
3. The accuracy of isodose plots and graphic displays;
4. The accuracy of the software used to determine sealed source positions from radiographic images; and
5. The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

Notes

Ariz. Admin. Code § R9-7-743
New Section R9-7-743 recodified from R12-1-743 at 24 A.A.R. 813, effective March 22, 2018 (Supp. 18-1).