Ariz. Admin. Code § R9-7-743 - Therapy-related Computer Systems
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing shall include, as applicable, verification of:
1. The source-specific input
parameters required by the dose calculation algorithm;
2. The accuracy of dose, dwell time, and
treatment time calculations at representative points;
3. The accuracy of isodose plots and graphic
displays;
4. The accuracy of the
software used to determine sealed source positions from radiographic images;
and
5. The accuracy of electronic
transfer of the treatment delivery parameters to the treatment delivery unit
from the treatment planning system.