(a) The user shall ensure fluoroscopic
equipment meets the following:
(1) The tube
housing shall be of diagnostic type.
(2) The target-to-panel or target-to-table
top distance should not be less than 18 inches and shall not be less than 12
inches.
(3) The total filtration
permanently in the useful beam shall not be less than 2.5 millimeters aluminum
equivalent. This requirement may be assumed to have been met if the half-value
layer is not less than 2.5 millimeters aluminum at normal operating
voltages.
(4) The equipment shall
be so constructed that the entire cross-section of the useful beam is
attenuated by a primary protective barrier. This barrier is usually the viewing
device, either a conventional fluoroscopic screen or an image intensification
mechanism.
(A) The lead equivalent of the
barrier of conventional fluoroscopes shall be at least 1.5 millimeters for
equipment capable of operating up to 100 kVp, at least 1.8 millimeters for
equipment whose maximum operating potential is greater than 100 kVp and less
than 125 kVp, and at least 2.0 millimeters for equipment whose maximum
operating potential is 125 kVp or greater. Special attention must be paid to
the shielding of image intensifiers so that neither the useful beam nor
scattered radiation from the intensifier can produce a radiation hazard to the
operator or personnel. With the fluorescent screen 14 inches (35 cm) from the
panel or table top, the exposure rate 2 inches (5 cm) beyond the viewing
surface of the screen shall not exceed 30 mR/hr for each R per minute at the
table top with the screen in the useful beam without a patient and with the
fluoroscope operating at the highest potential employed.
(B) Collimators shall be provided to restrict
the size of the useful beam to less than the area of the barrier. For
conventional fluoroscopes this requirement is met if, when the adjustable
diaphragm is opened to its fullest extent, an unilluminated margin is left at
all edges of the fluorescent screen with the screen centered in the beam at a
distance of 35 cm (14 inches) from the panel or table top.
For image intensified fluoroscopy, shutters shall be
provided which can be adjusted to restrict the X-ray field to the visible
portion of the image receptor during fluoroscopy. For systems employing
rectangular X-ray fields and circular image receptors, this requirement is met
if the collimated beam forms a square which circumscribes, and is tangent to,
the circular margin of the image receptor.
(C) The tube mounting and the carrier shall
be so linked together that the carrier always intercepts the entire useful
beam. The X-ray exposure shall automatically terminate when the carrier is
removed from the useful beam.
(D)
Collimators and adjustable diaphragms or shutters to restrict the size of the
useful beam shall provide the same degree of protection as is required of the
tube housing.
(5) The
exposure switch shall be of the dead-man type.
(6) Each fluoroscopic unit shall be equipped
with a manual-reset cumulative timing device, activated by the exposure switch,
which will either indicate elapsed exposure time by a signal audible to the
fluorocopist or turn off the apparatus when the total exposure exceeds a
predetermined limit not exceeding five minutes in one or a series of
exposures.
(7) Useful beam exposure
rate.
(A) All fluoroscopic equipment. For
routine fluoroscopy, the exposure rate measured at the point where the center
of the useful beam enters a typical patient shall be as low as is practicable
and shall not exceed 5 roentgens per minute under the conditions specified
herein. This limit shall not apply during magnification procedures or the
recording of fluoroscopic images where higher exposure rates are required.
Compliance with this paragraph shall be determined using the measuring
specifications of subsection (a)(7)(D), plus the following procedures when the
automatic exposure rate control is used:
1.
The useful beam exposure rate shall be measured with a phantom equivalent to 9
inches of water or 7 7/8 inches of lucite, intercepting the entire useful
beam.
2. If the X-ray source is
below the table, the X-ray exposure rate shall be measured with the nearest
part of the imaging assembly located at 14 inches above the table
top.
3. The field size at the point
of exposure rate measurement shall be at least 6 1/4 square inches in area in
the plane perpendicular to the central ray.
(B) Fluoroscopic equipment manufactured after
August 1, 1974, and equipped with automatic exposure rate controls.
Fluoroscopic equipment which is provided with automatic exposure rate control
shall not be operable at any combination of tube potential and current which
will result in an exposure rate in excess of 10 roentgens per minute at the
point where the center of the useful beam enters the patient, except during
recording of fluoroscopicimages, or when an optional high level control is
provided. When so provided, the equipment shall not be operable at any
combination of tube potential and current which will result in an exposure rate
in excess of 5 roentgens per minute at the point where the center of the useful
beam enters the patient unless the high level control is activated. Special
means of activation of high level controls shall be required. The high level
control shall only be operable when continuous manual activation is provided by
the operator. A continuous signal audible to the fluoroscopist shall indicate
that the high level control is being employed.
(C) Fluoroscopic equipment manufactured after
August 1, 1974, without automatic exposure rate controls. Fluoroscopic
equipment which is not provided with automatic exposure rate control shall not
be operable at any combination of tube potential and current which will result
in an exposure rate in excess of 5 roentgens per minute at the point where the
center of the useful beam enters the patient, except during recording of
fluoroscopic images, or when an optional high level control is activated.
Special means of activation of high level controls shall be required. The high
level control shall only be operable when continuous manual activation is
provided by the operator. A continuous signal audible to the fluoroscopist
shall indicate that the high level control is being employed.
(D) Measuring useful beam exposure rate
compliance.
1. If the X-ray tube is below the
table, the exposure rate shall be measured 1 centimeter above the tabletop or
cradle.
2. If the X-ray tube is
above the table, the exposure rate shall be measured at 30 centimeters above
the tabletop with the end of the beam-limiting device or spacer positioned as
closely as possible to the point of measurement.
3. In a C-arm type of fluoroscope, the
exposure rate shall be measured at 30 centimeters from the input surface of the
fluoroscopic imaging assembly.
(8) Mobile fluoroscopic equipment shall meet
the requirements of this section where applicable, except that:
(A) Inherent provisions shall be made so that
the machine is not operated at a source-skin distance of less than 30 cm (12
inches).
(B) Image intensification
shall always be provided. Conventional fluoroscopic screens shall not be
used.
(C) It shall be impossible to
operate a machine when the collimating cone or diaphragm is not in
place.
(D) It shall be impossible
to energize the useful beam of a mobile fluoroscope unless the entire useful
beam is intercepted by the image receptor.
(9) Devices which indicate the X-ray tube
potential and current shall be provided, and should be located in such a manner
that the operator may monitor the tube potential and current during
fluoroscopy.
(10) A shielding
device of at least 0.25 millimeters lead equivalent shall be provided for
covering the bucky-slot during fluoroscopy.
(11) Whenever practicable, protective drapes,
or hinged or sliding panels, of at least 0.25 millimeters lead equivalent shall
be provided between the patient and the fluoroscopist to intercept scattered
radiation which would otherwise reach the fluoroscopist and others near the
machine. Such devices shall not substitute for wearing of a protective
apron.
(b) The user shall
ensure the following:
(1) Protective aprons
of at least 0.25 mm lead equivalent shall be worn in the fluoroscopy room by
each person, except the patient, whose body is likely to be exposed to 5 mR/hr
or more.
(2) On fluoroscopes with
automatic exposure controls the operator shall monitor the tube current and
potential at least once each week to ascertain that they are in their usual
ranges for a given set of operating parameters. This requirement may be met by
adjusting the controls to usual settings for fluoroscoping an average patient,
and using a phantom of any suitable material with attenuation roughly
equivalent to six to ten inches of water. Whenever the monitored tube current
or potential vary in a way which could increase the patient X-ray exposure rate
by more than 25% over the latest exposure rate measurement required by
subsection (b)(3), the cause(s) for the change shall be determined promptly and
the patient exposure rate shall be remeasured. On fluoroscopes with manual
exposure control only, the operator shall monitor the tube current and
potential at least once each day during use to ascertain that they are within
the normal ranges used by the facility. A written log shall be kept of all
monitored readings and shall include at least the tube current and potential,
the date, identification of the fluoroscope, and name of the person who did the
monitoring. Records of all monitored readings shall be preserved at the
facility for at least three years.
(3) Measurements of the table top or patient
exposure rate shall be made at least once each year for units with automatic
exposure control, and at least once each 3 years for units without automatic
exposure control, and immediately following alteration or replacement of a
major component, such as the X-ray tube, the exposure controls, the imaging
assembly, and the power source.
(4)
On the cineradiography equipment, the exposure rates to which patients are
normally subjected shall be determined at least once each year, and immediately
following alterations or replacement of a major component, such as the X-ray
tube, the exposure controls, the imaging assembly, and the power
source.
(c) The user
shall record for each patient the cumulative air kerma, if provided by the
equipment, or, if it is not provided, the total fluoroscopic irradiation time.
The terms "cumulative air kerma" and "fluoroscopic irradiation time" are as
defined in title 21, Code of Federal Regulations, Part 1020.30(b) referenced in
section
30305(a)(4). This
record shall be maintained for three years and made available for inspection by
the Department. The recorded value shall be trackable to the particular patient
but need not be retained in the patient's medical record.
Notes
Cal. Code
Regs. Tit. 17, §
30307
1.
Renumbering and amendment filed 9-4-73 as an emergency; effective upon filing
(Register 73, No. 36). Approved by State Building Standards Commission
11-30-73.
2. Certificate of Compliance filed 12-28-73 (Register 73,
No. 52).
3. Amendment filed 6-24-80; effective thirtieth day
thereafter (Register 80, No. 26).
4. Amendment filed 6-18-87;
operative 7-18-87 (Register 87, No. 28).
5. Amendment of subsections
(a), (a)(4), (a)(7)(A), (b) and (b)(2) and amendment of NOTE filed 5-13-2020;
operative 10-1-2020 pursuant to Government Code section 11343.4(b)(2) (Register
2020, No. 20).
6. New subsection (c) filed 8-18-2020; operative
10-1-2020 (Register 2020, No. 34).
Note: Authority cited: Sections 114975, 115000 and
131200, Health and Safety Code. Reference: Sections 114965, 114970, 115000,
115060, 131050, 131051 and 131052, Health and Safety
Code.
1. Renumbering
and amendment filed 9-4-73 as an emergency; effective upon filing (Register 73,
No. 36). Approved by State Building Standards Commission 11-30-73.
2. Certificate of Compliance filed 12-28-73 (Register 73, No. 52).
3. Amendment filed 6-24-80; effective thirtieth day thereafter (Register 80,
No. 26).
4. Amendment filed 6-18-87; operative 7-18-87 (Register 87,
No. 28).
5. Amendment of subsections (a), (a)(4), (a)(7)(A), (b) and
(b)(2) and amendment of Note filed 5-13-2020; operative
10/1/2020
pursuant to Government Code section 11343.4(b)(2) (Register
2020, No. 20).
6. New subsection (c) filed 8-18-2020;
operative 10/1/2020 (Register
2020, No. 34).