Cal. Code Regs. Tit. 17, § 30312 - Therapeutic X-Ray Installations
(a) The user shall ensure therapeutic X-ray
equipment meets the following:
(1) The tube
housing shall be of therapeutic type.
(2) For equipment installed on or before
August 1, 1979, permanent diaphragms or cones used for collimating the useful
beam shall afford the same degree of protection as the housing. Adjustable or
removable beam-defining diaphragms or cones shall transmit not more than 5
percent of the useful beam obtained at the maximum kilovoltage and with maximum
treatment filter.
(3) For equipment
installed after August 1, 1979, permanent beam-defining devices or diaphragms
shall afford the same degree of protection as the housing. Adjustable or
interchangeable beam-defining devices shall transmit no more than 2 percent of
the useful beam for the portion of the useful beam which is to be attenuated by
the beam limiting device. Measurements shall be averaged over an area up to but
not exceeding 100 square centimeters at the normal treatment
distance.
(4) Filters shall be
secured in place to prevent them from dropping out during treatment. A filter
indication system shall be used on all therapy machines using interchangeable
filters. It shall indicate, from the control panel, or from the control
station, the presence or absence of any filter except compensating filters, and
it shall be designed to permit easy identification of the filter in place. The
filter slot shall be so constructed that the radiation escaping through it does
not exceed 1 roentgen per hour at 1 meter, or, if the patient is likely to be
exposed to radiation escaping from the slot, 30 roentgens per hour at 5
centimeters from the external opening. Each interchangeable filter shall be
marked with its thickness and material.
(5) The X-ray tube shall be so mounted that
it cannot turn or slide with respect to the aperture.
(6) Means shall be provided to immobilize the
tube housing during stationary portal treatment.
(7) A suitable exposure control device such
as an automatic timer, exposure meter, or dose meter shall be provided to
terminate the exposure after a preset time interval or preset exposure or dose
limit. A timer shall be provided to terminate the exposure after a preset time
regardless of what other exposure limiting devices are present. Means shall be
provided for the operator to terminate the exposure at any time.
(8) Equipment utilizing shutters to control
the useful beam shall have a shutter position indicator on the
control.
(9) An easily discernible
indicator which shows whether or not X-rays are being produced shall be on the
control panel.
(10) Mechanical
and/or electrical stops shall be provided on X-ray machines capable of
operating at 150 kVp or above to insure that the useful beam is oriented only
toward primary barriers.
(11) When
the relationship between the beam interceptor (when present) and the useful
beam is not permanently fixed, mechanical or electrical stops shall be provided
to insure that the beam is oriented only toward primary
barriers.
(b) The user
shall ensure the following:
(1) When a patient
must be held in position for radiation therapy, patient immobolization devices
shall be used.
(2) No patient other
than the one being treated shall be in the treatment room during
exposure.
(3) No person other than
the patient shall be in the treatment room when the tube is operated at
potentials exceeding 150 kVp. At operating potentials of 150 kVp or below,
persons other than the patient and operator may be in the treatment room for
good reason but only if they are adequately protected and their radiation
exposure is appropriately monitored.
(4) A calibration of the output of each
therapeutic X-ray system shall be performed before the system is first used for
irradiation of a patient, and thereafter at intervals not to exceed 24 months.
Therapeutic X-ray equipment shall not be used for any therapy treatments except
at those combinations of effective energy, field size, and treatment distance
for which the equipment has been calibrated. The calibration shall be performed
by or under the direct supervision of a TCP. After any change which might
significantly alter the output, spatial distribution, or other characteristics
of the therapy beam, the parameters which might be affected shall be measured.
(A) For therapy systems operating at
potentials above 500 kVp, the determinations included in the calibration shall
be provided in sufficient detail so that the absorbed dose in tissue in the
useful beam may be calculated to within 5 percent. The calibration shall
include, but shall not be limited to, the following determinations:
1. Verification that the equipment is
operating in compliance with the design specifications concerning the light
localizer, the side light and back-pointer alignment with the isocenter when
these specifications are known and applicable, variation in the axis of
rotation for the table, gantry and jaw system, and beam flatness and symmetry
at specified depths.
2. The
relative dose at various depths in a tissue equivalent phantom for each
effective energy and the ranges of field sizes and treatment distances used for
radiation therapy.
3. The
congruence between the radiation field and the field indicated by the
localizing device.
4. The
uniformity of the radiation field and its dependency upon the direction of the
useful beam.
5. The absolute dose
per unit time and dose per monitor setting.
(B) For therapy systems operating at
potentials between 150 kVp and 500 kVp inclusive, the calibration shall
include, but shall not be limited to, the following determinations:
1. The exposure rates and/or dose rates for
each combination of field size, technique factors, filter, and treatment
distance used.
2. The degree of
congruence between the radiation field and the field indicated by the
localizing device if such device is present.
3. An evaluation of the uniformity of the
radiation field symmetry for the field sizes used, and any dependence upon tube
housing assembly orientation.
(5) All new installations and existing
installations not previously surveyed shall have a radiation protection survey
performed by or under the supervision of a TSP or a TCP. If the survey shows
that supplementary shielding is required, a resurvey shall be performed by or
under the supervision of a TSP or a TCP after its installation. In addition, a
resurvey shall be made after every change which might decrease radiation
protection significantly. The surveyor shall report his findings in writing to
the user. The report shall indicate whether or not the installation is in
compliance with all applicable radiation protection requirements of this
section. The user shall report the findings of the survey in writing to the
Department within 15 days of his receipt of the survey report.
(6) The exposure rate or dose rate of the
useful beam and the size and shape of the useful beam shall be known with
reasonable certainty at all times during operation of the therapeutic X-ray
systems for medical purposes.
(7)
Spot checks shall be performed at least once each week for therapeutic X-ray
systems operating at potentials above 500 kVp, and at least once each month for
therapeutic X-ray systems operating at 500 kVp or below.
(A) The measurements taken during spot checks
shall demonstrate the degree of consistency of the operating characteristics
which can affect the radiation output of the system or the radiation delivered
to a patient during a therapy procedure.
(B) For systems in which the calibrating
person believes beam quality can vary significantly, spot checks shall include
beam quality checks.
(C) The spot
check procedures shall be in writing and shall have been developed or approved
by a TCP. The written spot check procedures shall specify when measurements and
determinations indicate an inconsistency or potential change in radiation
output. When more than the minimum frequency of spot checking is necessary, the
spot check procedures shall specify the frequency at which spot checks are to
be performed.
(D) When spot check
results are erratic or inconsistent with calibration data, a TCP shall be
consulted immediately and the cause of the inconsistency corrected before the
system is used for patient irradiation.
(8) Calibration of the therapy beam shall be
performed with a measurement instrument which has been calibrated within the
preceding two years directly, or through no more than one exchange, at the
National Institute of Standards and Technology, or facility determined
acceptable by the Department. In addition, indirect spot checks or
intercomparisons of measurement instruments with secondary standards shall be
made at least each six months.
(9)
Reports of each radiation safety survey spot check and calibration performed
pursuant to this section shall be maintained at the facility for at least three
years. A copy of the treatment data developed from the latest calibration shall
be available for use by the operator at the treatment control
station.
Notes
2. Certificate of Compliance filed 12-28-73 (Register 73, No. 52).
3. Amendment of subsection (c)(5) filed 12-12-75; effective thirtieth day thereafter (Register 75, No. 50).
4. Amendment filed 6-24-80; effective thirtieth day thereafter (Register 80, No. 26).
5. Amendment filed 6-18-87; operative 7-18-87 (Register 87, No. 28).
6. Change without regulatory effect amending subsection (b)(7)(C) and (b)(8) filed 11-1-91 pursuant to section 100, title 1, California Code of Regulations (Register 92, No. 5).
7. Amendment of subsections (a), (b)-(b)(1), (b)(4), (b)(5)-(7) and (b)(7)(C)-(D) and amendment of NOTE filed 5-13-2020; operative 10-1-2020 pursuant to Government Code section 11343.4(b)(2) (Register 2020, No. 20).
Note: Authority cited: Sections 114975, 115000 and 131200, Health and Safety Code. Reference: Sections 114965, 114970, 115000, 115060, 131050, 131051 and 131052, Health and Safety Code.
2. Certificate of Compliance filed 12-28-73 (Register 73, No. 52).
3. Amendment of subsection (c)(5) filed 12-12-75; effective thirtieth day thereafter (Register 75, No. 50).
4. Amendment filed 6-24-80; effective thirtieth day thereafter (Register 80, No. 26).
5. Amendment filed 6-18-87; operative 7-18-87 (Register 87, No. 28).
6. Change without regulatory effect amending subsection (b)(7)(C) and (b)(8) filed 11-1-91 pursuant to section 100, title 1, California Code of Regulations (Register 92, No. 5).
7. Amendment of subsections (a), (b)-(b)(1), (b)(4), (b)(5)-(7) and (b)(7)(C)-(D) and amendment of Note filed 5-13-2020; operative