Cal. Code Regs. Tit. 17, § 30337 - Requirements for Use of Cabinet X-ray Systems
(a) As used in this section:
(1) "Access panel" means any barrier or panel
which is designed to be removed or opened for maintenance or service purposes,
requires tools to open, and permits access to the interior of the
cabinet;
(2) "Aperture" means any
opening in the outside surface of the cabinet, other than a port, which remains
open during generation of X-rays;
(3) "Door" means any barrier which is
designed to be movable or opened for routine operation purposes, does not
generally require tools to open, and permits access to the interior of the
cabinet;
(4) "External surface"
means the outside surface of the radiation machine, including the high-voltage
generator, doors, access panels, latches, control knobs, and other permanently
mounted hardware and including the plane across any aperture or port;
(5) "Ground fault" means an accidental
electrical grounding of an electrical conductor;
(6) "Port" means any opening in the outside
surface of the radiation machine which is designed to remain open, during
generation of X-rays, for the purpose of conveying material to be irradiated
into and out of the cabinet, or for partial insertion for irradiation of an
object whose dimensions do not permit complete insertion into the
cabinet;
(7) "Primary beam" means
the radiation emitted directly from the target and passing through the window
of the X-ray tube;
(8) "Safety
interlock" means a device, which is intended to prevent the generation of
radiation when access by any part of the human body to the interior of the
detection system through a door or access panel is possible;
(9) X-ray system means an assemblage of
components for the controlled generation of X-rays;
(10) "X-ray tube" means any electron tube,
which is designed for the conversion of electrical energy into X-ray
energy.
(b) Cabinet X-ray
systems shall meet and be continually maintained to ensure the following are
met:
(1) Radiation emitted from the system
shall not, under any condition of use, exceed an exposure of 0.5 milliroentgen
in one hour at any point five centimeters (cm) outside the external surface, or
any door or port. The exposure shall be determined by measurements averaged
over a cross-sectional area of ten square cm with no linear dimension greater
than five cm with doors and access panels fully closed as well as fixed at any
position, which will allow the generation of X-rays;
(2) The insertion of any part of the human
body through any port into the primary beam shall not be possible. The
insertion of any part of the human body through any aperture shall not be
possible;
(3) The system shall have
a lock-and-key control, which will ensure that X-ray generation is not possible
with the key removed. When the system is not in use, the key shall be removed
and controlled to prohibit unauthorized use of the system;
(4) The system shall have a control or
controls to initiate and terminate the generation of X-rays other than by
functioning of a safety interlock or the main power control;
(5) The system shall have two independent
means (indicators), which indicate when and only when X-rays are being
generated. At least one of the indicators shall be illuminated when X-rays are
being generated. One, but not both, of the required indicators may be a
milliammeter labeled to indicate X-ray tube current. All other indicators shall
be legibly labeled "X-RAY ON." If the X-ray generation period is less than
one-half second, the indicators shall be activated for one-half second and
shall be discernible from any point at which initiation of X-ray generation is
possible. Failure of a single component of the system shall not cause failure
of both indicators to perform their intended function. The system shall have
additional means other than milliammeters as needed to insure that at least one
indicator is visible from each door, access panel, and port. If the X-ray
generation period is less than one-half second, the indicators shall be
activated for one-half second and be legibly labeled "X-RAY ON";
(6) In systems used to inspect objects such
as, but not limited to, baggage, boxes, backpacks, purses, and mail, the system
shall be designed such that:
(A) During an
exposure or preset succession of exposures of one-half second or greater
duration, the operator can terminate the exposure or preset succession of
exposures at any time and is in a position that permits surveillance of the
ports and doors during X-ray generation; and
(B) During an exposure or preset succession
of exposures of less than one-half second duration, completion of the exposure
in progress may continue but shall enable the operator to prevent additional
exposures;
(7) There
shall be permanently affixed or inscribed:
(A) At the location of any controls which can
be used to initiate X-ray generation, a clearly legible and visible label
bearing the statement: "Caution: X-Rays Produced When Energized;" and
(B) Adjacent to each port a clearly legible
and visible label bearing the statement: "Caution: Do Not Insert Any Part of
the Body When System is Energized--X-ray Hazard;"
(8) Each door shall have a minimum of two
safety interlocks. One, but not both of the required interlocks shall be such
that door opening results in physical disconnection of the energy supply
circuit to the high-voltage generator, and such disconnection shall not be
dependent upon any moving part other than the door;
(9) Each access panel shall have at least one
safety interlock;
(10) Following
interruption of X-ray generation by the functioning of any safety interlock,
use of a control provided in accordance with subsection (b)(4) shall be
necessary for resumption of X-ray generation;
(11) Failure of any single component of the
system shall not cause failure of more than one required safety interlock;
and
(12) A ground fault shall not
result in the generation of X-rays.
(c) A user shall not allow any individual to
operate a cabinet X-ray system until such individual has:
(1) Received copies of, instruction in, and
demonstrated understanding of, the user's operating and emergency procedures by
obtaining a passing grade of at least 80 percent on a written examination
covering this material. The written examination shall be at least 50 questions
in length; and
(2) Demonstrated
competence to use the radiation machines by obtaining a passing grade of at
least 80 percent on a practical examination covering this material. The
practical examination shall be at least 25 questions in length. An individual
operating such a system need not obtain radiographer
certification.
(d)
Interlocks shall be annually tested to ensure they function as
designed.
(e) The user shall
conduct an annual evaluation of the cabinet X-ray system to ensure compliance
with title 10, Code of Federal Regulations, Part 20, subpart D incorporated by
reference in section
30253.
(f) Individuals shall not be exposed to the
primary beam.
(g) Documentation
demonstrating compliance with this section shall be maintained for three years
and kept available for inspection.
Notes
2. Certificate of Compliance filed 2-22-74 (Register 74, No. 8).
3. Change without regulatory effect adding NOTE (Register 87, No. 11).
4. Amendment of section heading, section and NOTE and filed 4-11-2008; operative 5-11-2008 (Register 2008, No. 15).
Note: Authority cited: Sections 114975, 115000, 131050, 131051 and 131200, Health and Safety Code. Reference: Sections 114965, 114970 and 115060, Health and Safety Code.
2. Certificate of Compliance filed 2-22-74 (Register 74, No. 8).
3. Change without regulatory effect adding NOTE (Register 87, No. 11).
4. Amendment of section heading, section and Note and filed 4-11-2008; operative 5-11-2008 (Register 2008, No. 15).
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.