Cal. Code Regs. Tit. 22, § 73677 - Autoclaving, Sterilizing and Disinfecting
(a) All facilities shall make provision
within the facility, or by acceptable arrangements for outside service, for
proper sterilization of dressings, utensils, instruments and solutions used in
the facility.
(b) All sterilization
of supplies and equipment in the facility shall be under the supervision of a
person qualified by experience and training.
(1) Autoclaves and sterilizers where provided
shall be maintained in operating condition at all times.
(2) Instructions for operating autoclaves and
sterilizers shall be posted.
(3)
Written instructions for preparing supplies for autoclaving and for the
operation of autoclaves shall be available to personnel. This shall include the
method of packaging, loading the autoclave, temperature and pressure to be
applied and the period of exposure.
(4) Autoclaves shall be checked at least
weekly to verify that recording thermometers and indicating thermometers
reasonably coincide.
Records of recording thermometers shall be checked for each load to verify that the performance of the autoclave is in accordance with established standards for sterilization. Records of recording thermometers shall be preserved for one year. Recording thermometers are not required on portable sterilizers and autoclaves.
(5) Autoclaving results shall be checked at
least monthly by bacteriological tests. Records of results of such tests shall
be preserved at least one year.
(6)
Sterilized supplies and equipment shall be dated.