6 CCR 1010-22-8 - INSTRUMENTS/STERILIZATION
8-801 Instrument
Cleaning
(a) All instruments that penetrate
body tissue shall be properly cleaned prior to packaging and sterilization. All
other instruments shall be cleaned and disinfected after each use.
(b) All instruments placed in the procedure
area shall be repackaged and re sterilized.
(c) Employees shall wear heavy-duty,
multi-use, and waterproof gloves while cleaning instruments.
(d) Used instruments shall be soaked in a
disinfectant until cleaning can be performed. The solution shall be changed in
a time as recommended by the solution manufacturer.
(e) Instruments shall be disassembled for
cleaning.
(f) All instrument
components shall be cleaned, either manually or in an ultrasonic cleaner, using
the appropriate cleaning agent specific to the type of cleaning
performed.
8-802
Instrument packaging/wrapping
(a) Employees
shall wear clean gloves while packaging/wrapping instruments.
(b) Instruments shall be wrapped or packaged
with a sterilizer indicator on or in each package.
(c) All packages shall be labeled with the
time and date of sterilization. Packages will no longer be considered sterile
six months after the date of sterilization.
8-803 Instrument Sterilization
(a) The sterilizer shall be designed and
labeled as a medical instrument sterilizer.
(b) The operators' manual for the sterilizer
shall be available on the premise and the sterilizer shall be operated
according to manufacturer's recommendations.
(c) The sterilizer shall be cleaned and
maintained according to manufacturer's specifications.
(d) A sterilizer load log shall be maintained
for a minimum of three years at the facility and made available for inspection.
The log shall contain the following documentation for each load:
1. Description of instruments contained in
the load;
2. Date of sterilization
load, and time or other unique identifier if more than one load is processed
during a single day;
3. Sterilizer
cycle time and temperature;
4.
Indication of proper sterilization of instruments, as evidenced by the
appropriate color indicator change on each package. Indicator used shall be
compatible with the sterilization process being used; and
5. Action taken when appropriate color
indicator change did not occur.
(e) Sterilizer Monitoring
1. Sterilizer monitoring shall be performed
at least monthly (unless more frequent monitoring is specified by the
manufacturer) by using a commercial biological monitoring (spore)
system.
2. All biological
indicators shall be analyzed by a laboratory independent from the
establishment.
3. Biological
indicator test results shall be maintained on the premises for a minimum of
three (3) years and must be available for inspection at all
times.
8-804
Instrument Storage
(a) Hands shall be washed
prior to handling sterilized instrument packs.
(b) After sterilization, the instruments
shall be stored in a dry, clean area reserved for storage of sterile
instruments.
8-805 Single
Use Items
(a) Single use items shall not be
used on more than one client and shall be disposed of after the
procedure.
(b) Contaminated single
use needles, razors and other sharps shall be disposed of immediately in
approved sharps containers.