Haw. Code R. § 11-114-7 - Accuracy tests
(a) Every accuracy
test procedure shall be approved by the DUI coordinator in writing and shall
include, but not be limited to the following requirements:
(1) The test shall be conducted by a
supervisor;
(2) At least two
different reference samples and an air blank shall be run with each accuracy
test;
(3) Reference samples shall
be chosen so that their target values are not less than 0.04gm alcohol /210
liters and not greater than 0.25gm alcohol /210 liters;
(4) Reference sample target values shall
differ from each other by at least 0.04gm alcohol /210 liters;
(5) Reference sample test results which vary
from the target value by more than plus or minus 0.01gm alcohol /210 liters or
plus or minus ten percent, whichever is greater, shall be cause for the breath
alcohol testing instrument used to be removed from service until the fault has
been corrected; and
(6) An accuracy
test shall be performed on an operating instrument at intervals not to exceed
thirty-one days.
(b)
Operating instruments do not include instruments in storage or being repaired.
Upon return to operation from storage or repair an accuracy test shall be
performed.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.