Ill. Admin. Code tit. 32, § 380.30 - Labeling
a) A label bearing
the words "Caution - Radiation - This Equipment Produces Radiation When
Energized" shall be placed near any switch which energizes a tube.
b) A sign bearing the words "Caution - High
Intensity X-Ray Beam", or words having a similar intent, shall be placed in the
area immediately adjacent to each tube head. The sign shall be so located that
it is clearly visible to any person operating, aligning, or adjusting the unit,
or handling or changing a sample.
c) A clearly visible indication of the
presence of an x-ray beam shall be provided on or immediately adjacent to each
tube head.
d) A clearly visible
indication of the status (i.e., open or closed) of each shutter shall be
provided.