Iowa Admin. Code r. 481-943.13 - Disinfecting and sterilizing instruments and equipment
All
(1)
Disinfection .
a. Nonporous tools and implements.
(1) Immersion method. After each use, all immersible
nonporous tools and implements shall be disinfected by cleaning the tools and
implements followed by complete immersion in a disinfectant . Disinfectant
solutions shall be mixed according to manufacturer label instructions. The
manufacturer's listed contact time for effectively eliminating all pathogens
shall be adhered to at all times.
(2) Nonimmersion method. After each use, any nonporous
item that cannot be immersed in a disinfectant shall be cleaned with soap or
detergent and water to remove all organic material and then sprayed or wiped
with disinfectant . Minimum disinfectant contact time as listed on the
manufacturer's label shall be followed. Nonimmersible tools and implements
include but are not limited to scissors, trimmers, clippers, handles of hair
dryers and curling/flat irons.
b. Disinfected implements shall be stored in a
disinfected, dry, covered container and shall be isolated from contaminants.
Such container shall be disinfected at least once each week and whenever
visibly dirty.
c. Disinfectant
solutions shall be changed as instructed on the solution's manufacturer label
or whenever visibly dirty.
d.
Electric file bits.
(1) After each use, all
visible debris shall be removed from diamond, carbide, natural and metal bits
by cleaning with either an ultrasonic cleaner or immersion of each bit in
acetone for five to ten minutes.
(2) After they are cleaned, diamond, carbide, natural
and metal bits shall be disinfected by complete immersion in an appropriate
disinfectant . Minimum disinfectant contact time as listed on the manufacturer's
label shall be followed.
(2)
Sterilization . Ultraviolet (UV)
light boxes are prohibited and are not an acceptable method of sterilization .
a. Tools and implements may be sterilized by
one of the following methods:
(1) Steam
sterilizer, registered and listed with the FDA and used according to the
manufacturer's instructions. If steam sterilization , or moist heat, is
utilized, heat exposure shall be at a minimum of 121°C/250°F for at
least 30 minutes;
(2) Dry heat
sterilizer, registered and listed with the FDA and used according to the
manufacturer's instructions. If dry heat sterilization is utilized, heat
exposure shall be at a minimum of 171°C/340°F for at least 60
minutes;
(3) Autoclave
sterilization equipment, calibrated to ensure that it reaches the temperature
required by the manufacturer's instructions. If autoclave sterilization
equipment is utilized, spore testing by a contracted independent laboratory
shall be performed at least every 30 days. If a positive spore test is
received, the autoclave may not be used until a negative spore test is
received. The establishment must maintain a log of each autoclave use, all
testing samples and results, and a maintenance log of all maintenance performed
on the device. Maintenance shall be performed according to the manufacturer's
instructions. The establishment must have available for inspection the
autoclave maintenance log for the most recent 12 months; or
(4) Chemical sterilization with a hospital grade
liquid which, if used, shall be used according to the directions on the label.
When chemical sterilization is used, items shall be fully submerged for at
least ten minutes.
b.
Sterilization equipment shall be maintained in working order. The equipment
shall be checked at least monthly and calibrated to ensure that it reaches the
temperature required by the manufacturer's instructions.
This rule is intended to implement Iowa Code section 157.6.