Kan. Admin. Regs. § 28-35-242b - General requirements for all diagnostic X-ray systems
In addition to meeting the other requirements of this part, each diagnostic X-ray system shall meet the following requirements:
(a) Warning label. The control panel
containing the main power switch shall bear this or an equivalent warning
statement, which shall be legible and accessible to view: "WARNING: This X-ray
unit could be dangerous to patient and operator unless safe exposure factors
and operating instructions are observed."
(b) Battery charge indicator. On each
battery-powered X-ray generator, a visual means shall be provided on the
control panel to indicate whether the battery is in a state of charge for
proper operation.
(c) Leakage
radiation from the diagnostic source assembly. The leakage radiation from the
diagnostic source assembly measured at a distance of one meter in any direction
from the source shall not exceed 25.8 µC/kg (100 milliroentgens) in one hour
when the X-ray tube is operated at the leakage technique factors specified by
the manfacturer. Compliance shall be determined by measuring the leakage
radiation averaged over an area of 100 square centimeters, with no linear
dimension greater than 20 centimeters.
(d) Radiation from components other than the
diagnostic source assembly. The radiation emitted by a component other than the
diagnostic source assembly shall not exceed 0.5 µC/kg (2 milliroentgens) in one
hour at five centimeters from an accessible surface of the component in an
assembled X-ray system operated under any design conditions. Compliance shall
be determined by measuring the radiation averaged over an area of 100 square
centimeters, with no linear dimension greater than 20 centimeters.
(e) Beam quality.
(1) Half-value layer.
(A) The half-value layer of a given X-ray
tube potential shall not be less than the values shown in table I in this
paragraph. Linear interpolation and extrapolation may be used if necessary to
determine the half-value layer at an X-ray tube potential that is not listed in
table I.
Table I
Operating range (kilovolts peak) Measured potential (kilovolts peak) Half-value layer (millimeters of aluminum)
Below 50 30 0.3
40 0.4
49 0.5
50 through 70 50 1.2
60 1.3
70 1.5
Above 70 71 2.1
80 2.3
90 2.5
100 2.7
110 3.0
120 3.2
130 3.5
140 3.8
150 4.1
(B) For capacitor energy storage equipment,
compliance shall be determined with the system fully charged and set at 10 mAs
for each exposure.
(C) The
required minimal half-value layer of the useful beam shall include the
filtration contributed by all materials that are permanently located between
the source and the patient.
(2) Filtration controls. For each X-ray
system that has variable kVp and variable filtration for the useful beam, a
device shall link the kVp selector with the filter or filters and shall prevent
an exposure, unless the minimum amount of filtration necessary to produce the
required HVL is in the useful beam for the given kVp that has been selected.
(f) Multiple tubes. If
two or more radiographic tubes are controlled by one exposure switch, each tube
that has been selected shall be clearly indicated before initiation of the
exposure. This indication shall be both on the X-ray control panel and at or
near the tube housing assembly that has been selected.
(g) Mechanical support of the tube head. The
tube housing assembly supports shall be adjusted so that the tube housing
assembly remains stable during each exposure, unless tube housing movement is a
designed function of the X-ray system.
(h) Technique indicators.
(1) The technique factors to be used during
each exposure shall be indicated before the exposure begins. If automatic
exposure controls are used, the technique factors that are set before the
exposure shall be indicated.
(2)
The requirements of paragraph (h)(1) may be met by permanent marking on
equipment that has fixed technique factors. The indication of technique factors
shall be visible from the operator's position, except in the case of spot films
made by the fluoroscopist.
(i) Maintaining compliance. All diagnostic
X-ray systems and their associated components used on humans and certified
pursuant to the federal X-ray equipment performance standard in 21 CFR part
1020 shall be maintained in compliance with the applicable requirements of that
standard.
(j) Locks. All
position-locking, holding, and centering devices that are on X-ray system
components and systems shall function as intended and shall be maintained
according to each manufacturer's recommendations.