A.
Obtaining
proficiency test samples
A laboratory must obtain proficiency test samples from an
approved provider meeting the criteria below:
1. Approved providers of proficiency test
samples
a. The National Environmental
Laboratory Accreditation Program (NELAP) Institute, also known as The NELAP
Institute (TNI);
b. EPA;
or
c. Other proficiency test
providers that are also accreditation bodies.
2. Criteria for approval
The accreditation officer must approve a proficiency testing
provider, if the proficiency testing provider:
a. Is compliant with a national or
international standard and is a recognized proficiency testing provider
acceptable for use by TNI;
b.
Clearly defines the scope of each proficiency testing study;
c. Evaluates results from all proficiency
testing studies using the acceptance criteria described in national or
international standards or those specified by the accreditation
officer;
d. Scores each result as
either "acceptable," "not acceptable," "satisfactory," "unsatisfactory,"
"pass," "fail," "no evaluation" or "not reported;"
e. Provides to participant laboratories
reports that include:
(i) The provider name
in the header;
(ii) The laboratory
name, laboratory address (physical location), EPA laboratory ID number in the
header, and the name, title and telephone number of the laboratory point of
contact in the header or cover letter;
(iii) The study number and study type in the
header;
(iv) The shipment date and
closing date of the study in the header;
(v) The date of any amended report, if
applicable, in the header;
(vi)
Analyte name for each analyte included in the sample;
(vii) Method description;
(viii) Laboratory value as
reported;
(ix) Assigned values and
acceptance values reported to three significant figures, except for tests
requiring reports of presence or absence of the analyte;
(x) The acceptable/not acceptable
status;
(xi) A "no evaluation"
score for reported values containing alpha characters;
(xii) An indication of the length of the
report presented by either "page X of Y" or the total number of pages with each
page numbered consecutively; and
(xiii) For amended reports, an indication of
the amended results, including a brief description of the reason for the
amendment.
f. Sends
reports of results no later than 21 calendar days after the study closing
date;
g. Maintains the overall
effectiveness of the provider's quality system to indicate that samples
provided for testing are verifiable, homogeneous and stable;
h. Makes available to the accreditation
officer and any participating laboratory, upon request, a complete report of
the provider's analytical data and documentation of the provider's quality
system, which relates to the assigned values, homogeneity and stability of a
particular proficiency testing study;
i. Makes available to the accreditation
officer, upon request, a report listing the total number of participating
laboratories and the number of laboratories scoring "not acceptable" for each
analyte;
j. Supplies reports to the
accreditation officer in an electronic format acceptable to the accreditation
officer; and
k. Supplies the
laboratory with a proficiency test sample formulated from a lot that has not
been previously sent to the laboratory. If the lot was previously used in a
proficiency test sample or its assigned values sent to any laboratory, the
original proficiency test sample tracking ID must be obliterated, and the new
sample tracking ID must be unique.
B. Availability of proficiency test samples
1. The accreditation officer must determine
that a proficiency test sample for a particular field of testing is not
available if:
a. None of the approved
providers list the proficiency test sample through published catalogs, websites
or other widely distributed literature; or
b. None of the approved providers make the
proficiency test sample available in a form similar to routine samples. For
example, proficiency test samples may be considered unavailable if the
preparation instructions require the laboratory to perform pretreatment steps
not normally associated with the requirements of the approved methods. In this
context, dilution of the proficiency test sample is not considered
pretreatment.
2. If the
accreditation officer determines that no approved provider has proficiency test
samples for a field of testing, the accreditation officer must request written
documentation from the laboratory of QC data, meeting the minimum requirements
under this rule to evaluate the capability of the laboratory to perform
testing.
C.
Proficiency testing requirements to obtain or maintain
accreditation
1. A laboratory seeking
to obtain accreditation must successfully complete at least one proficiency
test sample, unless a proficiency test is not available, for each requested
field of testing, no more than six months before the date the laboratory
submits its application.
2. When a
laboratory is granted accreditation status, it must continue to complete
proficiency testing studies for each field of testing each calendar year and
maintain a history of at least one acceptable evaluation for each field of
testing out of the most recent two proficiency test sample results submitted to
the proficiency test provider.
3.
When a laboratory has attained accreditation and requests to add a field of
testing to its scope of accreditation, the laboratory must submit acceptable
proficiency testing results for that field of testing, analyzed no more than
six months before the date that the laboratory submits its
application.
4. To maintain
accreditation, a laboratory must complete the annual study, and any corrective
action study required, by October 31.
5. If the laboratory is accredited by the
MLAP through equivalency, and maintains accreditation or certification by
another approving body, proficiency test sample results for all analytes and
programs must be sent at the request of the accreditation officer.
D.
Proficiency test sample
analysis requirements
1. A laboratory
must analyze proficiency test samples in the same manner as used for routine
environmental samples, using the same staff, sample tracking, sample
preparation and analysis methods, SOPs, calibration techniques, QC procedures
and acceptance criteria. The laboratory must follow sample preparation steps
for the proficiency test sample as instructed by the approved proficiency test
provider for which the proficiency test sample was obtained.
2. Laboratories under the same ownership are
not to participate in the same study by the same approved proficiency test
provider for the same fields of testing, except when a study is not again
available for that field of testing by any approved proficiency test provider
within the calendar year.
3. Prior
to the closing date of a study, laboratory personnel, including corporate
personnel, may not:
a. Communicate with any
individual at another laboratory concerning the analysis of the proficiency
test sample prior to the closing date of the study;
b. Subcontract the analysis of any
proficiency test sample or a portion of a proficiency test sample to another
laboratory for any analysis;
c.
Knowingly receive and analyze any proficiency test sample or portion of a
proficiency test sample from another laboratory for which the results of the
proficiency test sample are intended for use for initial or continued
accreditation; or
d. Attempt to
obtain the assigned value of any proficiency test sample.
E.
Proficiency test sample
reporting requirements
The laboratory must evaluate and report the analytical result
for accreditation as follows:
1. For
instrument technology that employs a multi-point calibration, the working range
of the calibration under which the proficiency test sample is analyzed must be
the same range as used for routine samples.
a.
A result for any proficiency test at a concentration above or equal to the
lowest calibration standard must be reported as the resultant value.
b. A result for any proficiency test at a
concentration less than the lowest calibration standard must be reported as
less than the value of the lowest calibration standard.
c. A result for any proficiency test greater
than the highest calibration standard must be diluted to fall within the range
of the calibration curve.
2. For instrument technology (e.g., ICP-AES
or ICP-MS) that employs standardization with a zero point and a single point
calibration standard, the laboratory must evaluate the analytical result in the
same range as used for routine samples.
a. A
result for any proficiency test at a concentration above or equal to the RL
must be reported as the resultant value.
b. A result for any proficiency test at a
concentration less than the RL must be reported as less than the value of the
RL.
c. A result for any proficiency
test greater than the high calibration standard must be diluted to be within
the working range.
3. A
laboratory must ensure that the proficiency test results include the correct
physical address of the laboratory.
4. The laboratory must report the analytical
results for accreditation to the proficiency test provider on or before the
closing date of the study using the reporting format specified by the
proficiency test provider.
5. On or
before the closing date of the study, the laboratory must authorize the
proficiency test provider to release the laboratory's final evaluation report
directly to the laboratory's primary approving body. The MLAP will evaluate
only results received directly from the proficiency test provider.
6. A laboratory must supply results by
authorizing the approved proficiency test provider to release all accreditation
and corrective action results to the accreditation officer by an electronic
format specified by the accreditation officer.
7. For programs other than the Drinking Water
Program, a laboratory may use one proficiency test sample to analyze and report
results for multiple methods by the same technology, provided the sample is
analyzed under analytical conditions which satisfy all technologies reported
and the most stringent method QC requirements are fulfilled.
8. Errors in reporting the proper matrix, the
method used or the tested analytes in the proficiency test study by the
laboratory must be graded as "not acceptable" by the MLAP.
9. A laboratory may not request a revised
report from the proficiency test provider when the revisions to the report are
due to any error on the part of the laboratory.
10. Laboratories also accredited under NELAP
must designate one of the two annual proficiency test studies performed to be
sent to the MLAP. Additional proficiency test studies received will be
evaluated and subject to the same corrective action requirements found in
SubSection
I of this
Section.
F.
Proficiency test sample study record retention
1. The laboratory must maintain copies of all
written, printed and electronic records pertaining to proficiency test sample
analyses for five years or for as long as is required by the applicable
regulatory program, whichever is greater.
2. Proficiency test records must include, but
not be limited to:
a. Bench sheets;
b. Instrument strip charts or
printouts;
c. Data
calculations;
d. Data reports;
and
e. Proficiency test study
report forms used by the laboratory to record proficiency test
results.
3. The
laboratory must make all retained records available to assessors during on-site
assessments of the laboratory.
G.
Evaluation of results
1. All study data released from the
proficiency test provider will be scored for compliance with this
rule.
2. A laboratory must
demonstrate acceptable performance, as determined by the approved provider, for
each field of testing reported.
3.
For the purpose of initial or continuing accreditation, the accreditation
officer will deem unacceptable any reported results not meeting the criteria
under this Section.
4. Proficiency
test samples analyzed or reported after the study closing date are not valid
for compliance with the proficiency testing requirements under this
Section.
H.
Questionable proficiency test samples
Upon notice from a laboratory and verification by the
approved provider that a proficiency test sample did not meet the requirements
in this Section, the accreditation officer may determine that the affected
laboratory must analyze another proficiency test sample for that field of
testing.
I.
Corrective
actions for unacceptable results
1.
When an approved provider notifies a laboratory that a proficiency test sample
result for any reported field of testing is unacceptable, the laboratory must:
a. Review each "not acceptable" result to
determine:
(i) Cause of the error;
and
(ii) Corrective action(s) the
laboratory need(s) to take to correct the problem.
b. Document the cause(s) and corrective
action(s) the laboratory has taken to correct each problem that caused or
contributed to the "not acceptable" result;
c. Submit the corrective action report to the
MLAP within 30 days of receiving the notification of unacceptable results from
the approved provider;
d. Request a
supplemental proficiency test study within 30 days of closing the corrective
action report;
e. Ensure that
results of all proficiency test samples are received by the accreditation
officer and submitted directly by the proficiency test provider; and
f. Notify the MLAP if the laboratory fails a
second proficiency test (PT) for an accredited parameter.
J.
Grounds for suspension of
accreditation
The following constitute grounds for suspension of
accreditation:
1. Failure to produce
acceptable results in two consecutive proficiency testing studies for the same
field of testing; or
2. Failure to
submit an acceptable corrective action report in response to an unacceptable
proficiency testing result.
K.
Requirements for reinstatement of
accreditation after suspension
1. The
following requirements must be met for reinstatement of accreditation following
a proficiency test related suspension:
a. The
laboratory must complete a corrective action report within 30 days of the
suspension;
b. The laboratory must
pass two unique, successive proficiency test studies, analyzed at least one day
apart; and
c. The proficiency tests
must be completed within 180 days of the suspension or before the expiration of
the current certificate, whichever is sooner, to restore
accreditation.
2. To
reinstate accreditation after suspension for failure to submit corrective
action, refer to Section
7(E).
L.
Additional samples for
compliance
The accreditation officer may require Maine accredited
laboratories to test additional proficiency test samples at any time to
determine compliance with this rule.