Md. Code Regs. 26.11.31.11 - Zero Alignment Audit
A.
General Requirements.
(1) A zero alignment
audit conducted in accordance with Regulation .08B of this chapter shall
compare the monitor responses to the simulated zero check and the actual clear
path condition. Primary and alternate methods for performing the zero alignment
audit are described in §A(2) and (3) of this regulation.
(2) Primary Zero Alignment Method.
(a) The primary zero alignment method shall
be performed under clear path conditions. This is accomplished for the
installed transmissometer where the process is not operating and the monitor
pathlength is free of particulate matter or the monitor may be removed from its
installation and set up under clear path conditions.
(b) No adjustment to the monitor should be
made other than the establishment of the proper monitor pathlength and correct
optical alignment of the transmissometer components. For some monitors it may
also be necessary to disable the zero compensation mechanism or to record the
amount of correction applied to the simulated zero condition.
(c) The monitor response to the clear path
condition and to the simulated zero condition shall be recorded as percent
opacity.
(d) The response
difference in percent opacity to the clear path and simulated zero conditions
shall be recorded as the zero alignment error. The simulated zero device shall
then be adjusted to provide the same response as a clear path
condition.
(e) The monitor shall
then be restored to its operating mode at the facility.
(3) Alternate Zero Alignment Method.
(a) An external, removable zero-jig may be
used to facilitate periodic checks of the simulated zero condition as an
alternate zero alignment audit method if:
(i)
The zero-jig setting is established for the specific monitor by comparison of
the monitor responses to the zero-jig and to clear path condition;
and
(ii) The zero-jig is capable of
producing a consistent zero response when it is repeatedly installed on the
monitor.
(b) The
zero-jig shall be protected when not in use to ensure that the setting
equivalent to zero opacity does not change.
(c) Source owners who use a zero-jig shall
perform a primary zero alignment audit and check of the zero-jig setting at
least once every 3 years.
B. Criteria for a Successful Audit.
(1) Performance Audit. Performance of the COM
shall be considered acceptable if it meets the following criteria:
(a) Stack Exit Correlation Error: < 2
percent opacity;
(b) Fault
Indicators: Inactive/no error messages present;
(c) Zero and Upscale Responses: < 2
percent opacity;
(d) Zero
compensation: < 4 percent opacity;
(e) Optical Alignment: misalignment error
< 2 percent opacity;
(f) Optical
Surface Dust Accumulation: < 4 percent opacity; and
(g) Calibration Error: < 3 percent
opacity.
(2) Zero
Alignment.
(a) Calculations for a zero
alignment audit shall be in accordance with the EPA document incorporated by
reference in Regulation .03B(2) of this chapter.
(b) The zero alignment is acceptable if the
error of the simulated zero check is less than 2 percent opacity before
adjustment.
(c) The simulated zero
check should be adjusted to provide the correct response each time the zero
alignment audit is performed.
(3) Out-of-Control Periods. The beginning of
the out-of-control period is the time corresponding to the completion of the
performance audit indicating unacceptable performance. The end of the
out-of-control period is the time corresponding to the completion of the
subsequent successful audit.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.