Mich. Admin. Code R. 333.5701 - Purpose and scope
Rule 701.
(1)
This part establishes requirements governing the use of computed tomography
(CT) scanners by, or on behalf of, a health practitioner licensed under article
15 of the act.
(2) This part
applies to all registrants who use a CT scanner for the intentional exposure of
humans for diagnostic imaging.
(3)
A CT scanner is exempt from this part if the scanner meets 1 of the following:
(a) Generates a peak power of 5 kilowatts or
less as certified by the manufacturer.
(b) Is used only for attenuation corrections
and anatomical markers as part of a positron emission tomography (PET/CT) or
single photon emission computed tomography (SPECT/CT) study.
(c) Is used as a simulator solely for
treatment planning purposes in conjunction with a megavoltage radiation therapy
unit.
(d) Is used solely for
intra-operative guidance tomography.
(4) In addition to the requirements of this
part, all registrants are subject to applicable parts of these rules and the
certificate of need review standards for computed tomography scanner
services.