Minn. R. 4740.2089 - STANDARDS, REAGENTS, AND BACTERIOLOGICAL MEDIA

A. Reference standards that are used in the laboratory must be obtained, when available, from the National Institute of Standards and Technology (NIST), manufacturers that supply NIST standards or NIST traceable standards, or an international standard-setting organization.
B. A laboratory must retain records for all standards, reagents, and bacteriological media. The records must include:
(1) identification of the manufacturer or vendor;
(2) certificate of analysis or purity, if supplied;
(3) lot number;
(4) date of receipt or preparation;
(5) preparer's initials, if applicable;
(6) method of preparation, when prepared in the laboratory;
(7) recommended storage conditions; and
(8) expiration date after which the material must not be used unless its reliability is verified by the laboratory.

Notes

Minn. R. 4740.2089
31 SR 446 44 SR 371

Statutory Authority: MS s 144.97; 144.98

State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.