Nev. Admin. Code § 459.174 - Duties of registrant
The registrant for an invasive intervention radiation machine shall:
1. Establish
and maintain policies and procedures to ensure the safe operation of each
invasive intervention radiation machine for which the Division has issued a
registration certificate to the registrant pursuant to NAC
459.156. The policies and
procedures established pursuant to this subsection must require that tests of
an invasive intervention radiation machine are performed in accordance with the
frequency required and consistent with the requirements of the manufacturer of
the radiation machine.
2. Ensure
that:
(a) The results of the tests required by
subsection 1 are analyzed to determine if there are any problems requiring
correction;
(b) The necessary
corrective action is taken whenever the results of a test for quality assurance
indicate that such action is required; and
(c) If necessary corrective action is taken,
the action is taken before the radiation machine is used to perform any
examination on a patient.
3. Ensure that a medical physicist who
satisfies the qualifications set forth in
21 C.F.R. §
900.12(a)(3) performs:
(a) Surveys in accordance with the frequency
and requirements described in
21 C.F.R. §
900.12(e)(9) and which
include the testing described in
21 C.F.R. §
900.12(e)(4)(iii) and
(e)(5)(xi); and
(b) The examinations required by section
10.
4. Ensure that
records of the tests required by subsection 1, the surveys required by
subsection 3 and the examinations required by section 10 are retained until the
Division authorizes disposal of the records.