N.J. Admin. Code § 7:18-9.5 - Requirements for acute toxicity testing samples
(a) Dilution water
samples for acute toxicity testing shall be collected, handled and preserved in
accordance with the following requirements:
1.
Dilution water is acceptable for use in a toxicity test only if healthy test
organisms survive in it through acclimation pursuant to
7:18-7.4(e)3 ii,
without showing any signs of stress, including but not limited to, abnormal
behavior, discoloration, infection or disease;
2. Dilution water samples shall either be
representative of the receiving water system that the effluent is discharged into,
or, as designated by the Department in the NJPDES permit, be an alternate or
reference water. Dilution water samples shall be collected in the following manner:
i. In non-tidal waters, dilution water samples
shall be collected from a location outside of the influence, but upstream of, the
effluent, except when the effluent is discharged into the headwaters of the water
body. Under those conditions the dilution water sample shall be obtained in
accordance with the procedures specified in (a)4 below;
ii. In estuarine waters, dilution water samples
shall be collected from a location outside of the influence of the effluent, except
when the effluent is discharged into the headwaters of the water body. Under those
conditions the dilution water sample shall be obtained in accordance with the
procedures specified in (a)4 below. Samples shall also be collected during the
outgoing tide up to and during low slack tide;
iii. In marine waters (that is, tidal saltwaters),
dilution water samples shall be collected from a location outside of the influence
of the effluent being tested;
iv. The
sampling location shall be such that the salinity of the sample shall be within the
salinity range for the receiving water immediately outside of the effluent mixing
zone;
v. When samples are collected from
streams or rivers, an integrated sample shall be collected. This is a sample that is
collected from bottom to the top of the water column so that the sample collected is
proportional to the flow. If only a grab sample can be taken it should be collected
at mid-depth in midstream;
vi. When
samples are collected from reservoirs or lakes, the effects of seasonal
stratification, runoff, and previous rain fall upon the chemical-physical
characteristics of the water shall be considered; and
vii. If the receiving water has a natural pH below
5.0 units, then the dilution water samples shall be adjusted to pH of 5.0 prior to
their use in test organism acclimation and/or toxicity testing.
3. If the receiving water is influenced by other
point sources of pollution so as to disqualify its use as dilution water in
accordance with the NJPDES permit, then the dilution water sample(s) shall be either
obtained from a location just above the other point sources in the case of streams,
or outside the zone of influence of other point sources in the case of other water
bodies;
4. If acceptable dilution water
cannot be obtained from the receiving water at any location because an effluent is
discharged into the receiving water headwaters, then some other unpolluted water,
meeting the following requirements, shall be used as an alternate in the following
order of preference:
i. Another surface water or
groundwater having a natural quality similar to that of the receiving water prior to
its pollution may be used; or
ii.
Reconstituted or artificial freshwater or saltwater having a natural quality similar
to that of the receiving water prior to its pollution may be used; and
iii. An alternate dilution water shall have a
total hardness, alkalinity, salinity, and specific conductance within 25 percent and
a pH within 0.4 units of the receiving water prior to its pollution, but not less
than a pH of 5.0 units;
5.
The preparation of reconstituted freshwater or saltwater, as an alternate dilution
water, shall comply with the following:
i.
Preparation of reconstituted freshwater shall be by the addition of reagent grade
chemicals to laboratory pure water as specified in SM16 p. 699-701, or EPA Acute
Methods #027F-1993; and
ii. Preparation
of a substitute or reconstituted saltwater dilution water shall either be through
the use of a hypersaline brine as specified in
7:18-7.4(b)8 ii, by
using commercial sea salts, or by the addition of reagent grade chemicals to
laboratory pure water as specified in SM16, p. 699-701 or EPA Acute Methods
#027F-1993.
6. Alteration of
dilution water samples shall be limited to the following:
i. Filtration through screening made of a
non-toxic material as specified in
7:18-7.3(a)1. This
screening shall have a mesh of two mm or larger for fishes or 0.45 microns or larger
for zooplankton and macrocrustaceans; and
ii. Adjustment of the salinity of dilution water
samples shall only be by either the addition of laboratory pure water to lower the
salinity or by the addition of either a hypersaline brine or artificial sea salts to
raise the salinity.
(1) Only a natural water
source, meeting the requirements for laboratory grade salt waters, shall be used to
produce a hypersaline brine; and
(2) A
hypersaline brine shall not exceed a salinity of 100 ppt;
7. Sample collection and transport
containers shall meet the requirements listed in
7:18-7.3(a)13. Prior to
sample collection all containers shall be rinsed with the dilution water and then
filled so that there should be little or no air in the container neck or
cap;
8. Dilution water sample storage
shall be in covered containers constructed of non-toxic materials as specified in
7:18-7.3(a)13;
and
9. Except for samples of laboratory
grade water being used as an alternate or reference dilution water as specified in
(a)4 above, samples shall not be stored for more than 150 hours and shall be
collected as close as possible to the time of use.
(b) Effluent samples for acute toxicity testing
shall be collected, handled, and preserved in accordance with the following
requirements:
1. The effluent sampling location
shall be the same as that specified in the NJPDES permit as the toxicity test
analysis sampling point unless otherwise specified by the Department. The Department
may specify an alternative sampling location when either of the following conditions
occur:
i. When there is better access to the
effluent at a point located between the final treatment and the discharge outfall.
That point shall be the sampling point; or
ii. When the chlorinated effluent is dechlorinated
prior to discharge, and the purpose of the test is to determine the toxicity levels
of the dechlorinated effluent. The sampling point shall be located after
dechlorination.
2. Samples
shall be representative of the discharge, taking into account the plant operating
conditions and the retention time of the effluent in the wastewater treatment
plant;
3. When performing flow-through
toxicity tests the following sampling procedures shall be adhered to in order to
insure a representative effluent sample:
i. If the
facility discharges continuously, the effluent shall be pumped directly and
continuously from the discharge line to the dilutor system for the duration of the
test; or
ii. If the facility discharges
continuously but the effluent cannot be pumped directly and continuously to the
dilutor system, then the following procedures shall be used:
(1) Twenty-four hour composite samples consisting
either of equal volumes taken once every hour, or flow-proportionate composite
sampling shall be collected and transported to the dilutor daily for the duration of
the test. Any surplus from the previous sample is to be discarded and the holding
container refilled with fresh effluent sample.
iii. If the facility discharges intermittently,
one of the following procedures shall be used:
(1)
When the effluent is discharged continuously only during a single work shift, or two
successive work shifts, at least one composite sample of sufficient volume to supply
the dilutor for 24 hours shall be collected daily during a single discharge period
for the duration of the test;
(2) When
the facility retains the wastewater during a work shift, then treats and releases it
in a batch discharge, a single grab sample of sufficient volume to supply the
dilutor for the intervening hours shall be collected and stored in accordance with
(b)10 below; and
(3) When the facility
discharges wastewater to an estuary during an outgoing tide, a single grab sample or
composite sample (as specified by the Department in the NJPDES permit), of
sufficient volume to set up the toxicity test shall be collected on the outgoing
tide. This procedure is repeated for the duration of flow-through toxicity
tests.
4. In order
to insure the collection of a representative effluent sample for a static or renewal
toxicity test, the following sampling procedures shall be followed:
i. If a static toxicity test is to be conducted,
effluent samples shall be collected only at the beginning of the test. If a renewal
toxicity test is to be conducted, then effluent samples shall be collected at the
beginning of the test and the test solutions renewed at least daily throughout the
duration of the test. Sampling for these renewal solutions shall comply with the
procedures specified in (b)4ii and iii below, and in (b)5 below;
ii. If the facility discharges wastewater
continuously the following procedures shall be used:
(1) Twenty-four hour composite samples consisting
of equal volumes collected at least once every hour or a flow proportionate 24 hour
composite sample shall be collected and used to set up a single toxicity test. This
procedure is repeated for the duration of renewal toxicity tests.
iii. If the facility discharges
wastewater intermittently one of the following procedures shall be used:
(1) When the effluent is discharged continuously
only during a single work shift, or two successive work shifts, at least one
composite sample, of sufficient volume to set up the toxicity test, shall be
collected. This procedure is repeated for the duration of renewal toxicity
test;
(2) When a facility retains the
wastewater during a work shift, then treats and releases it in a batch discharge, a
grab sample shall be collected during the discharge period. Sufficient volume of
sample shall be collected for the set up and renewal of the toxicity test during the
hours intervening between effluent discharges. Effluent samples shall be collected
and stored in accordance with (b)10 below; and
(3) When the facility discharges wastewater to an
estuary only during an outgoing tide, a single grab sample or composite sample (as
specified by the Department in the NJPDES permit), of sufficient volume to set up
the toxicity test shall be collected on the outgoing tide. This procedure is
repeated for the duration of renewal toxicity tests.
5. When the effluent to be sampled is a stormwater
discharge, the following sampling procedures shall be used for static, renewal, and
flow through toxicity tests:
i. The stormwater
discharge shall be a grab or composite sample either directly from the discharge
pipe during the precipitation event or from the retention pond during or immediately
after the precipitation event unless otherwise specified by the Department in the
NJPDES permit; and
ii. Sufficient sample
shall be collected during runoff from a precipitation event on the first day of
sampling to provide either for the set up and renewal, where applicable, or the
static or renewal toxicity test over its duration, or for the uninterrupted
operation of the dilutor system over the duration of the flow through toxicity test.
Samples shall be collected in this manner for each day the discharge persists during
the test period. Test sample renewal shall be conducted with the newest sample
available during the test period. Stormwater samples not used immediately shall be
stored in approved containers as specified in
7:18-7.3(a)14, and
preserved at 1.0 to 4.4 degrees Celsius. Samples shall not be stored for longer than
120 hours prior to use.
6.
Alteration of effluent samples shall be limited to:
i. Filtration through screening having a mesh of
two mm or larger;
ii. Introduction of
dry artificial sea salts or a hypersaline brine for the purpose of adjusting the
effluent test concentration salinity according to the procedures specified in
7:18-7.5(o);
iii. A laboratory may adjust an effluent sample
using a dechlorinating agent to reduce the level of chlorine in an effluent sample.
Since anhydrous sodium thiosulfate and other dechlorinating agents may contribute to
sample toxicity, the laboratory shall include an additional control containing the
dechlorinating agent in the acute toxicity test, in addition to the control chambers
specified in 7:18-7.5(b)5. The
amount of dechlorinating agent in the control shall be equal to that contained in
the highest effluent concentration tested. The laboratory shall document and report
adjustments and treatments of the effluent along with the test results. The
laboratory shall include in the documentation the type and amount of dechlorinating
agent which is added and the chlorine levels before and after
dechlorination.
7. Composite
or grab sample collection and handling containers shall meet the requirements listed
in 7:18-7.3(a)14. Prior to
sample collection, containers shall either be rinsed with the effluent or laboratory
pure water, as specified in
7:18-7.4(a), and then
filled so that there should be no air space in either the neck or the top of the
container;
8. Effluent samples shall be
stored in covered, sealed, containers constructed on non-toxic materials as
specified in 7:18-7.3(a)
14;
9. Unless the purpose of the
toxicity test is to ascertain the persistence of the toxic materials in an effluent,
testing shall begin within 36 hours of the collection of an effluent. For stormwater
discharge, the toxicity tests shall begin within 48 hours of collection conducted in
accordance with (b)5 above; and
10.
Samples that are collected for offsite testing shall be chilled during or
immediately after collection until adjustment to the test temperature prior to
initiating the test. When the sample arrives at the laboratory, the laboratory shall
log the sample in, measure the temperature of the sample, and record the temperature
on the chain-of-custody form and the raw data sheet. If samples are not immediately
prepared for testing, the laboratory shall store them between 1.0 and 4.4 degrees
Celsius until used.
(c) The
following chain-of-custody procedures shall be followed for effluent and dilution
water for all composite and grab samples in acute toxicity testing.
1. Only clean or new containers, as specified in
7:18-7.3(a)13 and 14,
previously rinsed with either laboratory pure water or the material being sampled,
shall be used for taking composite or grab samples;
2. Labels with an identification number shall be
affixed to all samples;
3. After a
sample has been collected, the appropriate information as to identity of the sample
shall be written on the label and the label affixed. The label shall remain affixed
until the test has begun and the surplus sample has been discarded;
4. Immediately upon delivery of a sample to the
certified environmental laboratory, the sample collector shall complete the
appropriate chain-of-custody section of the sample report form or chain-of-custody
form;
5. The chain-of-custody form shall
list at a minimum the following information:
i. The
sample number;
ii. The description of
samples;
iii. The specific location of
sample collection;
iv. The identity of
the person collecting the sample;
v. The
date and time of sample collection;
vi.
The date and time of custody transfer to laboratory (if the sample was collected by
a person other than laboratory personnel);
vii. The identity of the person accepting custody
(if the sample was collected by a person other than laboratory personnel);
viii. The date and time of initiation of
analyses;
ix. The identity of the person
performing analysis; and
x. The name and
identification number of environmental laboratory performing the analyses;
and
6. The laboratory
personnel accepting responsibility for the sample, as well as all other laboratory
personnel performing the analysis on that sample, shall sign the form containing the
chain-of-custody information.
Notes
See: 28 New Jersey Register 4098(a).
Amended by R.1997 d.192, effective
See: 28 New Jersey Register 4149(a), 29 New Jersey Register 2275(a).
In (b)9, changed time for toxicity testing from "24" to "36" hours; and substantially amended (b)10.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.