N.D. Admin Code 33.1-10-15-06 - Therapeutic radiation machines of less than five hundred kilovolts
1. Leakage
radiation. When the x-ray tube is operated at its maximum rated tube current
for the maximum kilovolts, the leakage air kerma rate shall not exceed the
value specified at the distance specified for that classification of
therapeutic radiation machine:
a. Five to
fifty kilovolts systems. The leakage air kerma rate measured at any position
five centimeters from the tube housing assembly shall not exceed one milligray
(100 mrad) in any one hour.
b.
Greater than fifty and less than five hundred kilovolts systems. The leakage
air kerma rate measured at a distance of one meter from the target in any
direction shall not exceed one centigray (1 rad) in any one hour. This air
kerma rate measurement may be averaged over areas no larger than one hundred
square centimeters. In addition, the air kerma rate at a distance of five
centimeters from the surface of the tube housing assembly shall not exceed
thirty centagray (30 rad) per hour.
c. For each therapeutic radiation machine,
the registrant shall determine, or obtain from the manufacturer, the leakage
radiation existing at the positions specified in subdivisions a and b for the
specified operating conditions. Records on leakage radiation measurements shall
be maintained at the installation for inspection by the department.
2. Permanent beam-limiting
devices. Permanent diaphragms or cones used for limiting the useful beam shall
provide at least the same degree of attenuation as required for the tube
housing assembly.
3. Adjustable or
removable beam-limiting devices.
a. All
adjustable or removable beam-limiting devices, diaphragms, cones, or blocks
shall not transmit more than five percent of the useful beam for the most
penetrating beam used; and
b. When
adjustable beam-limiting devices are used, the position and shape of the
radiation field shall be indicated by a light beam.
4. Filter system. The filter system shall be
so designed that:
a. Filters cannot be
accidentally displaced at any possible tube orientation;
b. An interlock system prevents irradiation
if the proper filter is not in place;
c. The air kerma rate escaping from the
filter slot shall not exceed one centigray (1 rad) per hour at one meter under
any operating conditions; and
d.
Each filter shall be marked as to its material of construction and its
thickness. For wedge filters, the nominal wedge angle shall appear on the wedge
or wedge tray, if permanently mounted to the tray.
5. Tube immobilization.
a. The x-ray tube shall be so mounted that it
cannot accidentally turn or slide with respect to the housing aperture;
and
b. The tube housing assembly
shall be capable of being immobilized for stationary portal
treatments.
6. Source
marking. The tube housing assembly shall be so marked that it is possible to
determine the location of the source to within five millimeters, and such
marking shall be readily accessible for use during calibration
procedures.
7. Beam block. Contact
therapy tube housing assemblies shall have a removable shield of material,
equivalent in attenuation to five-tenths millimeters of lead at one hundred
kilovolts, which can be positioned over the entire useful beam exit port during
periods when the beam is not in use.
8. Timer. A suitable irradiation control
device shall be provided to terminate the irradiation after a preset time
interval.
a. A timer with a display shall be
provided at the treatment control panel. The timer shall have a preset time
selector and an elapsed time or time remaining indicator;
b. The timer shall be a cumulative timer that
activates with an indication of "BEAM-ON" and retains its reading after
irradiation is interrupted or terminated. After irradiation is terminated and
before irradiation can be reinitiated, it shall be necessary to reset the
elapsed time indicator;
c. The
timer shall terminate irradiation when a preselected time has elapsed, if any
dose-monitoring system present has not previously terminated
irradiation;
d. The timer shall
permit accurate presetting and determination of exposure times as short as one
second;
e. The timer shall not
permit an exposure if set at zero;
f. The timer shall not activate until the
shutter is opened when irradiation is controlled by a shutter mechanism unless
calibration includes a timer error correction to compensate for mechanical lag;
and
g. Timer shall be accurate to
within one percent of the selected value or one second, whichever is
greater.
9. Control panel
functions. The control panel, in addition to the displays required by other
provisions in this section, shall have:
a. An
indication of whether electrical power is available at the control panel and if
activation of the x-ray tube is possible;
b. An indication of whether x-rays are being
produced;
c. A means for indicating
x-ray tube potential and current;
d. The means for terminating an exposure at
any time;
e. A locking device which
will prevent unauthorized use of the therapeutic radiation machine;
and
f. A positive display of
specific filters in the beam.
10. Multiple tubes. When a control panel may
energize more than one x-ray tube:
a. It
shall be possible to activate only one x-ray tube at any time;
b. There shall be an indication at the
control panel identifying which x-ray tube is activated; and
c. There shall be an indication at the tube
housing assembly when that tube is energized.
11. Target-to-skin distance (TSD). There
shall be a means of determining the central axis target-to-skin distance to
within one centimeter and of reproducing this measurement to within two
millimeters thereafter.
12.
Shutters. Unless it is possible to bring the x-ray output to the prescribed
exposure parameters within five seconds after the x-ray "ON" switch is
energized, the beam shall be attenuated by a shutter having a lead equivalency
not less than that of the tube housing assembly. In addition, after the unit is
at operating parameters, the shutter shall be controlled by the operator from
the control panel. An indication of shutter position shall appear at the
control panel.
13. Low filtration
x-ray tubes. Each therapeutic radiation machine equipped with a beryllium or
other low-filtration window shall be clearly labeled as such upon the tube
housing assembly and shall be provided with a permanent warning device on the
control panel that is activated when no additional filtration is present, to
indicate that the dose rate is very high.
14. Facility design requirements for
therapeutic radiation machines capable of operating in the range fifty
kilovolts to five hundred kilovolts. In addition to shielding adequate to meet
requirements of section
33.1-10-15-09, the treatment
room shall meet the following design requirements:
a. Aural communication. Provision shall be
made for continuous two-way aural communication between the patient and the
operator at the control panel.
b.
Viewing systems. Provision shall be made to permit continuous observation of
the patient during irradiation and the viewing system shall be so located that
the operator can observe the patient from the control panel. The therapeutic
radiation machine shall not be used for patient irradiation unless at least one
viewing system is operational.
15. Additional requirements. Treatment rooms
that contain a therapeutic radiation machine capable of operating above one
hundred fifty kilovolts shall meet the following additional requirements:
a. All protective barriers shall be fixed
except for entrance doors or beam interceptors;
b. The control panel shall be located outside
the treatment room or in a totally enclosed booth, which has a ceiling, inside
the room;
c. Interlocks shall be
provided such that all entrance doors, including doors to any interior booths,
shall be closed before treatment can be initiated or continued. If the
radiation beam is interrupted by any door opening, it shall not be possible to
restore the machine to operation without closing the door and reinitiating
irradiation by manual action at the control panel; and
d. When any door referred to in subdivision c
is opened while the x-ray tube is activated, the air kerma rate at a distance
of one meter from the source shall be reduced to less than one milligray (100
mrad) per hour.
16. Full
calibration measurements.
a. Full calibration
of a therapeutic radiation machine subject to this section shall be performed
by, or under the direct supervision of, a radiation therapy physicist:
(1) Before the first medical use following
installation or reinstallation of the therapeutic radiation machine;
(2) At intervals not exceeding one year;
and
(3) Before medical use under
the following conditions:
(a) Whenever
quality assurance check measurements indicate that the radiation output differs
by more than five percent from the value obtained at the last full calibration
and the difference cannot be reconciled; and
(b) Following any component replacement,
major repair, or modification of components that could significantly affect the
characteristics of the radiation beam.
(4) Notwithstanding the requirements of
paragraph 3:
(a) Full calibration of
therapeutic radiation machines with multi-energy capabilities is required only
for those modes or energies, or both, that are not within their acceptable
range; and
(b) If the repair,
replacement, or modification does not affect all energies, full calibration
shall be performed on the affected energy that is in most frequent clinical use
at the facility. The remaining energies may be validated with quality assurance
check procedures against the criteria in subparagraph a of paragraph 3.
b. To satisfy
the requirement of subdivision a, full calibration shall include all
measurements recommended for annual calibration by NCRP report 69, "dosimetry
of x-ray and gamma ray beams for radiation therapy in the energy range ten keV
to fifty MeV" (1981).
c. The
registrant shall maintain a record of each calibration for the duration of the
registration. The record shall include the date of the calibration; the
manufacturer's name, model number, and serial number for both the therapeutic
radiation machine and the x-ray tube; the model numbers and serial numbers of
the instruments used to calibrate the therapeutic radiation machine; and the
signature of the radiation therapy physicist responsible for performing the
calibration.
17. Periodic
quality assurance checks.
a. Periodic quality
assurance checks shall be performed on therapeutic radiation machines subject
to this section, which are capable of operation at greater than or equal to
fifty kilovolts.
b. To satisfy the
requirement of subdivision a, quality assurance checks shall meet the following
requirements:
(1) The registrant shall
perform quality assurance checks in accordance with written procedures
established by the radiation therapy physicist; and
(2) The quality assurance check procedures
shall specify the frequency at which tests or measurements are to be performed.
The quality assurance check procedures shall specify that the quality assurance
check shall be performed during the calibration specified in subdivision a of
subsection 16. The acceptable tolerance for each parameter measured in the
quality assurance check, when compared to the value for that parameter
determined in the calibration specified in subdivision a of subsection 16,
shall be stated.
c. The
cause for a parameter exceeding a tolerance set by the radiation therapy
physicist shall be investigated and corrected before the system is used for
patient irradiation.
d. Whenever a
quality assurance check indicates a significant change in the operating
characteristics of a system, as specified in the radiation therapy physicist's
quality assurance check procedures, the system shall be recalibrated as
required in subdivision a of subsection 16.
e. The registrant shall use the dosimetry
system described in subdivision b of subsection 3 of section
33.1-10-15-04 to make the
quality assurance check required in subdivision b.
f. The registrant shall have the radiation
therapy physicist review and sign the results of each radiation output quality
assurance check within one month of the date that the check was
performed.
g. The registrant shall
ensure that safety quality assurance checks of therapeutic radiation machines
subject to section 33.1-10-15-06 are performed at intervals not to exceed one
month.
h. Notwithstanding the
requirements of subdivisions f and g, the registrant shall ensure that no
therapeutic radiation machine is used to administer radiation to humans unless
the quality assurance checks required by subdivisions f and g have been
performed within the thirty-day period immediately prior to said
administration.
i. To satisfy the
requirement of subdivision g, safety quality assurance checks shall ensure
proper operation of:
(1) Electrical
interlocks at each external beam radiation therapy room entrance;
(2) The "BEAM-ON" and termination
switches;
(3) Beam condition
indicator lights on the access doors, control console, and in the radiation
therapy room;
(4) Viewing systems;
and
(5) If applicable, electrically
operated treatment room doors from inside and outside the treatment
room.
j. The registrant
shall maintain a record of each quality assurance check required by subdivision
g for three years. The record shall include the date of the quality assurance
check; the manufacturer's name, model number, and serial number of the
therapeutic radiation machine; the manufacturer's name, model number, and
serial number for the instruments used to measure the radiation output of the
therapeutic radiation machine; and the signature of the individual who
performed the periodic quality assurance check.
18. Operating procedures.
a. The therapeutic radiation machine shall
not be used for irradiation of patients unless the requirements of subsection
16 of section 33.1-10-15-06 and subsection 17 of this section have been
met;
b. Therapeutic radiation
machines shall not be left unattended unless secured pursuant to subdivision e
of subsection 9;
c. When a patient
must be held in position for radiation therapy, mechanical supporting or
restraining devices shall be used;
d. The tube housing assembly shall not be
held by an individual during operation unless the assembly is designed to
require such holding and the peak tube potential of the system does not exceed
fifty kilovolts. In such cases, the holder shall wear protective gloves and
apron of not less than five-tenths millimeters lead equivalency at one hundred
kilovolts;
e. A copy of the current
operating and emergency procedures shall be maintained at the therapeutic
radiation machine control console; and
f. No individual other than the patient shall
be in the treatment room during exposures from therapeutic radiation machines
operating above one hundred fifty kilovolts. At energies less than or equal to
one hundred fifty kilovolts, any individual, other than the patient, in the
treatment room shall be protected by a barrier sufficient to meet the
requirements of 33.1-10-04.2-01 [
10 CFR
20.1201] .
19. Possession of survey instruments. Each
facility location authorized to use a therapeutic radiation machine in
accordance with this section shall possess appropriately calibrated portable
monitoring equipment. As a minimum, such equipment shall include a portable
radiation measurement survey instrument capable of measuring dose rates over
the range ten microsievert (1 mrem) per hour to ten millisievert (1,000 mrem)
per hour. The survey instruments shall be operable and calibrated in accordance
with section
33.1-10-15-08.
Notes
General Authority: NDCC 23.1-03-02; S.L. 2017, ch. 199, § 1
Law Implemented: NDCC 23.1-03-03; S.L. 2017, ch. 199, § 18
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.