Ohio Admin. Code 3701:1-58-24 - Calibration of survey instruments
(A) A licensee shall calibrate the survey
instruments used to show compliance with this chapter and Chapter 3701:1-38 of
the Administrative Code before first use, annually, and following a repair that
affects the calibration. A licensee shall:
(1) Calibrate all scales with readings up to
ten millisievert, or
( one thousand millirem) per hour with a radiation source;
(2) Calibrate two separated
separate
readings on each scale or decade that will be used to show compliance; and
(3) Conspicuously note on the
instrument the date of calibration.
(B) A licensee may not use survey instruments
if the difference between the indicated exposure rate and the calculated
exposure rate is more than twenty per cent
percent.
(C) A licensee shall retain a record of each
survey instrument calibration in accordance with rule
3701:1-58-78
of the Administrative Code.