Ohio Admin. Code 3701:1-58-60 - Dosimetry equipment
(A) Except for low dose-rate remote
afterloader sources where the source output or activity is determined by the
manufacturer, a licensee shall have a calibrated dosimetry system available for
use. To satisfy this requirement, one of the following two conditions must be
met:
(1) The system must have been calibrated
using a system or source traceable to the national
institute of standards and technology
"National
Institute of Standards and Technology" (NIST) and published protocols
accepted by nationally recognized bodies; or by a calibration laboratory
accredited by the american association of physicists
in medicine
"American Association of Physicists
in Medicine" (AAPM). The calibration must have been performed within the
previous two years and after any servicing that may have affected system
calibration; or
(2) The system
must have been calibrated within the previous four years. Eighteen to thirty
months after that calibration, the system must have been intercompared with
another dosimetry system that was calibrated within the past twenty-four months
by NIST or by a calibration laboratory accredited by the AAPM. The results of
the intercomparison must indicate that the calibration factor of the licensee's
system had not changed by more than two percent
per cent . The
licensee may not use the intercomparison result to change the calibration
factor. When intercomparing dosimetry systems to be used for calibrating sealed
sources for therapeutic units, the licensee shall use a comparable unit with
beam attenuators or collimators, as applicable, and sources of the same
radionuclide as the source used at the licensee's facility.
(B) The licensee shall have a
dosimetry system available for use for spot-check output measurements, if
applicable. To satisfy this requirement, the system may be compared with a
system that has been calibrated in accordance with paragraph (A) of this rule.
This comparison must have been performed within the previous year and after
each servicing that may have affected system calibration. The spot-check system
may be the same system used to meet the requirement in paragraph (A) of this
rule.
(C) The licensee shall
retain a record of each calibration, intercomparison, and comparison in
accordance with rule
3701:1-58-93
of the Administrative Code.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 08/15/2005, 10/04/2010
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005, 10/5/10
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.