Ohio Admin. Code 3701:1-58-69 - Full-inspection servicing for teletherapy and gamma stereotactic radiosurgery units
(A) A licensee shall have each teletherapy
unit and gamma stereotactic radiosurgery unit fully inspected and serviced
during source replacement or at intervals not to
exceed five years, whichever comes first, to assure proper functioning
of the source exposure mechanism and other safety
components. The interval between each full-inspection servicing shall not
exceed five years for each teletherapy unit and shall not exceed seven years
for each gamma stereotactic radiosurgery unit .
(B) This inspection and servicing may only be
performed by persons specifically licensed to do so by the director, the United
States nuclear regulatory commission, or an agreement state.
(C) A licensee shall keep a record of the
inspection and servicing in accordance with rule
3701:1-58-100
of the Administrative Code.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 08/15/2005, 10/04/2010, 08/10/2015