Ohio Admin. Code 3701:1-58-70 - Therapy-related computer systems for remote afterloader units, teletherapy units, and gamma stereotactic radiosurgery units
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(A) The source-specific input parameters
required by the dose calculation algorithm;
(B) The accuracy of dose, dwell time, and
treatment time calculations at representative points;
(C) The accuracy of isodose plots and graphic
displays;
(D) The accuracy of the
software used to determine sealed source positions from radiographic images;
and
(E) The accuracy of electronic
transfer of the treatment delivery parameters to the treatment delivery unit
from the treatment planning system.
Notes
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 08/15/2005
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.