Or. Admin. Code § 331-915-0080 - Approved Sterilization Standards for Tattooing
(1) Needles must be single use, used on one
client, then properly disposed of in an approved sharps container defined under
OAR
331-915-0000.
(2) All non-sterilized or reusable
instruments that come in direct contact with a client's skin or are exposed to
blood or other potentially infectious materials must be cleaned and sterilized
before use on a client or re-use on another client.
(3) New gloves must be worn during any
sterilization procedure.
(4) The
cleaning and sterilization process listed in subsection (5) of this rule is not
required if single-use prepackaged sterilized instruments, obtained from
suppliers or manufacturers, are used.
(5) Approved cleaning and sterilization
process for nonsterilized or reusable instruments includes the following
ordered method after each use:
(a) Place
instruments in an ultrasonic cleaner filled with an appropriate ultrasonic
solution including, but not limited to, an enzymatic cleaner. The ultrasonic
unit must be used according to the manufacturer's instructions. The ultrasonic
unit must operate at 40 to 60 kilohertz. The ultrasonic cleaner must remain
covered when in use. A self-contained instrument washers used to decontaminate
instruments prior to sterilization may be used in place of an ultrasonic
cleaner and used according to manufacturer instructions.
(b) Remove instruments from the ultrasonic
unit or a self-contained instrument washer. Clean instruments by manually
brushing or swabbing visible foreign matter and rinsing the instruments with
warm water and an appropriate detergent solution to remove blood and other
potentially infectious materials;
(c) Remove instruments from the ultrasonic
unit or self-contained instrument washer. All instruments must be rinsed, air
dried, and individually packaged in sterilization pouches that include use of a
color change indicator strip to assure sufficient temperature during each
sterilization cycle. The date the sterilization was performed must be applied
to the sterilization pouch; OR instruments that are sterilized in an autoclave
that do not require packaging or use of a color change indicator strip must be
sterilized after use on a client, stored in a clean disinfected location, and
resterilized immediately before performing a service on a client.
(d) Instruments must be sterilized by using
an autoclave sterilizer, steam, chemical or gas, registered and listed with the
FDA;
(e) A steam sterilization
integrator must be used to monitor the essential conditions of steam
sterilization for each autoclaved cycle. Results must be recorded in a log book
for each sterilization cycle. Each steam sterilization integrator must indicate
the date the sterilization cycle took place. Steam sterilization integrators
must be kept for a minimum of sixty days; and
(f) After sterilization, the sterilized
instruments must be stored in individually packaged sterilization pouches that
include a color change indicator strip listed under (5)(c) of this rule and in
a dry, disinfected, closed cabinet or other tightly covered container reserved
for the storage of such instruments. The date the sterilization was performed
must be applied to the sterilization pouch while being stored.
(6) Use of a biological monitoring
system ("spore tests") must be done at least once a month, verified through an
independent laboratory, to assure all microorganisms have been destroyed and
sterilization achieved.
(7) The
ultrasonic unit listed in subsection (5)(a) of this rule must be used, cleaned,
and maintained in accordance with manufacturer's instructions and a copy of the
manufacturer's recommended procedures for the operation of the ultrasonic unit
must be kept on file at the body art facility.
(8) All sterilization pouches with color
change indicator strips listed in subsection (5)(c) of this rule must contain a
chemical/temperature and/or humidity sensitive tapes, strips or pellets for
monitoring each sterilization cycle.
(9) Sterilization pouches with color change
indicator strips listed in subsection (5)(c) of this rule and steam
sterilization integrators listed in (5)(e) of this rule must be available at
all times for inspection by the Office.
(10) Biological spore test results listed in
subsection (6) of this rule must be immediately available at all times for
inspection by the Office and kept at facility premises for a minimum of two
years.
(11) The autoclave listed in
subsection (5)(d) must be used, cleaned, and maintained in accordance with
manufacturer's instructions and a copy of the manufacturer's recommended
procedures for the operation of the autoclave must be kept on file at the body
art facility.
(12) The expiration
date for sterilized instruments is one year from the date of sterilization
unless the integrity of the package is compromised.
(13) Sterilized instruments may not be used
if the package integrity has been breached, is wet or stained, or the
expiration date has exceeded without first meeting the requirements listed in
subsection (5) of this rule.
(14)
All sterilized instruments used in tattooing procedures must remain stored in
sterile packages and in a dry, disinfected, closed cabinet or other
tightly-covered container reserved for the storage of such instruments until
just prior to the performance of a tattooing procedure.
(15) If a biological spore test listed in
subsection (6) of this rule, result is positive, a licensee must discontinue
the use of that sterilizer (autoclave) until it has been serviced and a
negative spore test has been recorded before putting the sterilizer back into
service. Until a negative spore test has been received, the licensee must:
(a) Use an alternative sterilizer
(autoclave); or
(b) Use only
single-use instruments.
(16) If a positive spore test has been
received, all instruments sterilized prior to the positive spore test must be
repackaged and sterilized pursuant to subsection (5) of this rule, before
use.
(17) Following the receipt of
a positive spore test, the licensee or facility must notify all clients in
writing stating a positive spore test had been received during the time they
received their tattoo procedure. The licensee or facility must provide the
client with the risks of having a tattoo procedure done with instruments that
may not have been sterilized pursuant to (5)(c) and (d) of this rule.
Notes
Statutory/Other Authority: 690.390 & 690.405
Statutes/Other Implemented: 676.615, 690.390 & 690.405
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.