67 Pa. Code § 157.32 - Calibration of measurement systems
(a)
Calibration. Requirements for calibration shall be as follows:
(1) The sound level measurement system shall
be calibrated and appropriately adjusted at one or more frequencies in the
range from 250 to 1,000 hertz at the beginning of each series of measurements
and at intervals of 5-15 minutes thereafter until it has been established that
the sound level measurement system has not drifted from its calibrated level.
Once this fact has been established, calibrations may be made at intervals of
once every hour. A significant drift shall be considered to have occurred if a
0.3 decibel or more excursion is noted from the systems predetermined reference
calibration level. In the case of systems using displays with whole decibel
increments, the operator may usually judge when the 0.3 decibel drift has been
met or exceeded.
(2) The sound
level measurement system shall be checked periodically by its manufacturer, a
representative of its manufacturer, or a person of equivalent special
competence to verify that its accuracy meets the design criteria of the
manufacturer.
(b)
Acoustical calibrator. An acoustical calibrator of the
microphone coupler type design for the sound level measurement system in use
shall be used to calibrate the sound level measurement system in accordance
with subsection (a) of this section. The calibrator shall meet or exceed the
accuracy requirements specified in section
5.4.1 of the American National
Standards Institute Standard Methods for Measurements of Sound Pressure Levels,
ANSI §1.13 -1971, for calibrators for
field type measurements.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.