Tenn. Comp. R. & Regs. 0400-20-07-.74 - PERIODIC SPOT-CHECKS FOR GAMMA STERIOTACTIC RADIOSURGERY UNITS
(1) A licensee authorized to use a gamma
stereotactic radiosurgery unit for medical use shall perform spot-checks of
each gamma stereotactic radiosurgery facility and on each unit:
(a) Monthly;
(b) At the beginning of each day of use;
and
(c) After each source
installation.
(2) A
licensee shall have the authorized medical physicist:
(a) Establish written procedures for
performing the spot-checks required in paragraph (1) of this rule;
and
(b) Review the results of each
spot-check required by paragraph (1) of this rule within 15 days of the check.
The authorized medical physicist need not actually perform the spot-check
measurements. The authorized medical physicist shall notify the licensee as
soon as possible, in writing, of the results of the spot check.
(3) To satisfy the requirements of
subparagraph (1)(a) of this rule, spot-checks must, at a minimum:
(a) Assure proper operation of:
1. Treatment table retraction mechanism,
using backup battery power or hydraulic backups with the unit off;
2. Helmet microswitches;
3. Emergency timing circuits; and
4. Stereotactic frames and localizing devices
(trunnions).
(b)
Determine:
1. The output for one typical set
of operating conditions measured with the dosimetry system described in
paragraph (2) of Rule 0400-20-07-.68;
2. The difference between the measurement
made in part 1 of this subparagraph and the anticipated output, expressed as a
percentage of the anticipated output (i.e., the value obtained at last full
calibration corrected mathematically for physical decay);
3. Source output against computer
calculation;
4. Timer accuracy and
linearity over the range of use;
5.
On-off error; and
6. Trunnion
centricity.
(4) To satisfy the requirements of
subparagraphs (1)(b) and (c) of this rule, spot-checks must assure proper
operation of:
(a) Electrical interlocks at
each gamma stereotactic radiosurgery room entrance;
(b) Source exposure indicator lights on the
gamma stereotactic radiosurgery unit, on the control console, and in the
facility;
(c) Viewing and intercom
systems;
(d) Timer
termination;
(e) Radiation monitors
used to indicate room exposures; and
(f) Emergency off buttons.
(5) A licensee shall arrange for
the repair of any system identified in paragraph (3) of this rule that is not
operating properly as soon as possible. .
(6) If the results of the checks required in
paragraph (4) of this rule indicate the malfunction of any system, a licensee
shall lock the control console in the off position and not use the unit except
as may be necessary to repair, replace, or check the malfunctioning
system.
(7) A licensee shall retain
a record of each check required by paragraphs (3) and (4) of this rule in
accordance with 0400-20-07-.107.
Notes
Authority: T.C.A. ยงยง 68-202-101 et seq., 68-202-201 et seq. and 4-5-201 et seq.
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.