Utah Admin. Code R307-170-5 - General Requirements
(1) Each source
required to operate a continuous monitoring system is subject to the
requirements of
40 CFR 60.13(d) through
(j), except as follows:
(a) When minimum emission data points are
collected by the continuous monitoring system as required in
40 CFR
60.13 or applicable subparts, quality
assurance calibration and maintenance activities shall not count against
monitor availability.
(b) A
monitor's unavailability due to calibration checks, zero and span checks, or
adjustments required in
40 CFR
60.13 or R307-170 will not be considered a
violation of R307-170.
(c) Monitor
unavailability due to continuous monitoring system breakdowns will not be
considered a monitor unavailability violation provided that the owner or
operator demonstrates that the malfunction was unavoidable and was repaired
expeditiously.
(d) To supplement
continuous monitor data, a source with minimum continuous monitoring system
data collection requirements may conduct applicable reference method tests
outlined in 40 CFR 60, Appendix A, or as directed in the source's applicable
Subpart of the New Source Performance Standards.
(2) Each source shall monitor and record all
emissions data during all phases of source operations, including start-ups,
shutdowns, and process malfunctions.
(3) Each source operating a continuous
emissions monitoring system for compliance determination shall document each
out-of-control period in the state electronic data report.
(4) Each continuous monitoring system subject
to R307-170 shall be installed, operated, maintained, and calibrated in
accordance with applicable performance specifications found in 40 CFR 60
Appendix B and Appendix F.
(5) Each
continuous emissions monitoring system shall be configured so that calibration
gas can be introduced at or as near to the probe inlet as possible. Each source
shall conduct daily calibration zero drift and span drift checks and cylinder
gas audits by flowing calibration gases at the probe inlet, or as near to the
probe inlet as possible. Daily calibration drift checks and quarterly cylinder
gas audit data shall be recorded by the continuous emissions monitoring system
electronically to a strip chart recorder, data logger, or data recording
devices.
(6) No person shall tamper
with a continuous monitoring system.
(7) Any source that constructs two or more
emission point sources that may interfere with visible emissions observations
shall install a continuous opacity monitor to show compliance with visible
emission limitations on each obstructed stack, duct or vent that has a visible
emission limitation.
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.