Utah Admin. Code R315-264-1054 - Standards: Pressure Relief Devices in Gas/Vapor Service
(a) Except during pressure releases, each
pressure relief device in gas/vapor service shall be operated with no
detectable emissions, as indicated by an instrument reading of less than 500
ppm above background, as measured by the method specified in Subsection
R315-264-1063(c).
(b)
(1) After each pressure release, the pressure
relief device shall be returned to a condition of no detectable emissions, as
indicated by an instrument reading of less than 500 ppm above background, as
soon as practicable, but no later than 5 calendar days after each pressure
release, except as provided in Section
R315-264-1059.
(2) No later than 5 calendar days
after the pressure release, the pressure relief device shall be monitored to
confirm the condition of no detectable emissions, as indicated by an instrument
reading of less than 500 ppm above background, as measured by the method
specified in Subsection
R315-264-1063(c).
(c) Any pressure relief
device that is equipped with a closed-vent system capable of capturing and
transporting leakage from the pressure relief device to a control device as
described in Section R315-264-264-1060 is exempt from the requirements of
Subsection R315-264-1054(a) and (b).
Notes
State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.
No prior version found.