010-18 Wyo. Code R. §§ 18-5 - Inspection and testing of weights and measures used commercially
(a) Weighing and
measuring devices shall be inspected and tested according to a risk based
schedule determined by the Director.
(b) Exception. The Department shall exercise
no authority over weights and measures used pursuant to a written agreement
between the parties using the weighing device.
(i) For a weighing device to be exempt from
this section, parties must agree in writing to the use of the weighing device.
Appendix A to these rules shows an example of a written agreement. Copies of
the signed and dated written agreements shall be retained by the seller for a
minimum of 2 years.
(c) Standards shall be inspected and tested
as follows:
(i) Standards in possession and
use by agencies of the State shall be inspected and tested
biennially.
(ii) Registered service
technicians shall have their standards inspected and tested biennially by the
Department, and will receive certification upon payment of the laboratory fees
established in these regulations.
(iii) All standards that are new, recently
purchased, or have been damaged must not be used prior to inspection, testing,
and approval by the Department.
(iv) Exceptions.
(A) Standards in possession and use by
service technicians and that have been inspected and tested biennially by
another recognized state laboratory or NIST accredited private laboratory are
acceptable for use in Wyoming and exempt from this section.
(B) Standards which are self-contained, such
as coal mine hopper weights, shall be inspected and tested by the Department at
a minimum of once every five (5) years.