Ohio Admin. Code 3701:1-58-48 - Calibration measurements of brachytherapy sources

(A) Before the first medical use of a brachytherapy source, a licensee shall have:
(1) Determined the source output or activity using a dosimetry system that meets the requirements of paragraph (A) of rule 3701:1-58-60 of the Administrative Code;
(2) Determined source positioning accuracy within applicators; and
(3) Used published protocols currently accepted by nationally recognized bodies to meet the requirements of paragraphs (A)(1) and (A)(2) of this rule.
(B) Instead of a licensee making its own measurements as required in paragraph (A) of this rule, the licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the american association of physicists in medicine "American Association of Physicists in Medicine" (AAPM) that are made in accordance with paragraph (A) of this rule.
(C) A licensee shall mathematically correct the outputs or activities determined in paragraph (A) of this rule for radioactive decay at intervals consistent with one percent radioactive decay.
(D) A licensee shall retain a record of each calibration in accordance with rule 3701:1-58-89 of the Administrative Code.

Notes

Ohio Admin. Code 3701:1-58-48
Five Year Review (FYR) Dates: 4/7/2022 and 04/07/2027
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 08/15/2005, 10/04/2010
Five Year Review (FYR) Dates: 06/12/2015 and 06/01/2020
Promulgated Under: 119.03
Statutory Authority: 3748.02, 3748.04
Rule Amplifies: 3748.04
Prior Effective Dates: 8/15/2005, 10/4/10

State regulations are updated quarterly; we currently have two versions available. Below is a comparison between our most recent version and the prior quarterly release. More comparison features will be added as we have more versions to compare.


No prior version found.