As used in this rule, "radiation-generating equipment" means
radiation-generating equipment used for dental, veterinary, or medical purpose
but does not include therapeutic radiation-generating equipment.
(A) The director may, upon application
thereof or upon his or her own initiative, grant a variance to the
requirements
obligations of rules in this chapter as he or she
determines is authorized by law, provided that the registrant shows to the
satisfaction of the director that there is good cause for the variance, and
that the variance will not result in any undue hazard or effect on the public
health and safety or environment. The terms, conditions, and expiration of the
variance shall
will be set forth in writing by the director. Failure
to comply with the terms of the variance may result in immediate revocation of
the variance.
(B)
Except as specified in rule 3701:1-66-17 of the
Administrative Code, no
An individual
shall be exposed
will not expose any individual or animal to the useful
beam
except a patient
unless ordered by a licensed practitioner acting within his
or her scope-of-practice for dental
or
, medical
radiologic procedures and unless such exposure has
been authorized by a licensed practitioner within his or her scope of practice.
This provision specifically prohibits deliberate exposure for the following
purposes:
or radiation therapy purposes.
Exposing an individual for training, demonstration or other purposes is
forbidden unless otherwise specified in rules promulgated under Chapter 4773.
or 3748. of the Revised Code.
(1) Exposure of an individual for
training, demonstration, or any other non-medical purpose; and
(2) Exposure of an individual for
the purpose of self-referred screening except as authorized by the department
in accordance with paragraph (A) of rule 3701:1-66-17 of the Administrative
Code.
(C) The
handler
shall
will assure:
(1)
Every individual who performs radiologic procedures on human beings holds the
appropriate radiologic license as required
obligated by
Chapter 3701-72 of the Administrative Code and Chapter 4715. of the Revised
Code.
(2) Every individual who is
licensed to perform radiologic procedures is adequately instructed in the
registrant's safe operating procedures and can demonstrate competency in the
safe use of the equipment.
(3) The
individual responsible for radiation protection (IRRP) is qualified as one of
the following:
(a) Ohio licensed to operate
radiation-generating equipment;
(b)
Dental assistant certified to operate dental radiation-generating
equipment;
(c) Registered
veterinary technician and trained to operate veterinary radiation-generating
equipment;
(d) Certified by the
American registry of radiologic technologists in a pathway involving ionizing
radiation or certified by the nuclear medicine technologist certification
board;
(e) A radiation expert as
defined in rule
3701:1-66-01 of the
Administrative Code;
(f) A health
physicist certified by the American board of health physics; or
(g) An associate's degree or higher in health
physics, radiologic science, nuclear medicine or nuclear engineering.
(D) Any
radiation-generating equipment that does not meet the provisions set forth in
this rule or any other applicable equipment requirements
obligations of Chapter 3701:1-66 of the Administrative
Code shall
will not be used to irradiate patients unless the
director or a radiation expert determines that the non-compliance will not pose
a radiation risk and arrangements have been made to promptly correct the
non-compliance.
(E)
Radiation-generating equipment shall
will bear a warning label on the control panel, by the
exposure switch or by the main power switch which cautions individuals that
radiation is produced when it is energized.
(F) Unless otherwise specified in this
paragraph, radiation-generating equipment
shall
will meet the
following standards:
(1) On battery-powered
x-ray generators, visual means shall
will be provided on the control panel to indicate
whether the battery is in a state of charge adequate for proper
operation;
(2) The leakage
radiation from the diagnostic source assembly measured at a distance of one
meter in any direction from the source shall
will not exceed
0.88 milligray air kerma (one hundred milliroentgen exposure) in one hour when
the x-ray tube is operated at its leakage technique factors. Compliance
shall
will be
determined by measurements averaged over an area of one hundred square
centimeters with no linear dimension greater than twenty centimeters;
(3) Except for mammographic
radiation-generating equipment, the half-value layer (HVL) of the useful beam
for a given x-ray tube potential
shall
will not be less than the values shown in table 1. If
it is necessary to determine such HVL at an x-ray tube potential which is not
listed in table 1, linear interpolation or extrapolation may be made;
Table 1.
|
X-Ray Tube Voltage (kilovolt peak)
|
Minimum HVL (millimeter of aluminum)
|
|
Designed Operating Range
|
Measured Operating Potential
|
Specified Dental
Systems1
|
I- Other X-Ray
Systems2
|
II - Other X-Ray
Systems3
|
|
Below 51
|
30
|
1.5
|
0.3
|
0.3
|
|
40
|
1.5
|
0.4
|
0.4
|
|
50
|
1.5
|
0.5
|
0.5
|
|
51 to 70
|
51
|
1.5
|
1.2
|
1.3
|
|
60
|
1.5
|
1.3
|
1.5
|
|
70
|
1.5
|
1.5
|
1.8
|
|
Above 70
|
71
|
2.1
|
2.1
|
2.5
|
|
80
|
2.3
|
2.3
|
2.9
|
|
90
|
2.5
|
2.5
|
3.2
|
|
100
|
2.7
|
2.7
|
3.6
|
|
110
|
3.0
|
3.0
|
3.9
|
|
120
|
3.2
|
3.2
|
4.3
|
|
130
|
3.5
|
3.5
|
4.7
|
|
140
|
3.8
|
3.8
|
5.0
|
|
150
|
4.1
|
4.1
|
5.4
|
|
1 Dental x-ray systems
designed for use with intraoral image receptors and manufactured after December
1, 1980.
|
|
2 Dental x-ray systems
designed for use with intraoral image receptors and manufactured before or on
December 1, 1980, and all other x-ray systems subject to this section and
manufactured before June 10, 2006.
|
|
3 All x-ray systems, except
dental x-ray systems designed for use with intraoral image receptors, subject
to this section and manufactured on or after June 10, 2006.
|
(a) For capacitor
energy storage equipment, compliance with the requirements
obligations of this paragraph
shall
will be
determined with the system fully charged and a setting of ten
milliampere-seconds (mAs) for each exposure; and
(b) The required
obligated
minimal HVL of the useful beam shall
will include the filtration contributed by all
materials which are permanently between the source and the patient;
(4) For x-ray systems which have
variable kilovolt peak (kVp) setting and variable filtration for the useful
beam, a device shall
will link the kVp selector with the filter and
shall
will
prevent an exposure unless the minimum amount of filtration necessary to
produce the HVL required
obligated by paragraph (F)(3) of this rule is in the
useful beam for the given kVp which has been selected;
(5) Where two or more x-ray tubes are
controlled by one exposure switch, the tube that has been selected
shall
will be
clearly indicated prior to initiation of the exposure. This indication
shall
will be
both on the x-ray control panel and for dental equipment at or near the
selected tube housing assembly;
(6)
The x-ray tube housing assembly supports shall
will be adjusted
such that the tube housing assembly will remain stable during an exposure
unless tube housing movement is a designed function of the radiation-generating
equipment;
(7) The technique
factors to be used during an exposure shall
will be
indicated before the exposure begins. If automatic exposure controls are used,
the technique factors which are set prior to the exposure
shall
will be
indicated. This requirement
obligation
may be met by permanent markings on equipment having fixed technique factors.
Indication of technique factors shall
will be visible from the operator's position except in
the case of spot films taken during fluoroscopy procedures or dental intraoral
or panoral films; and
(8) All
position locking, holding, and centering devices on radiation-generating
equipment components shall
will function as designed by the
manufacturer.
(G) In
addition to other applicable radiation safety rules in Chapter 3701:1-66 of the
Administrative Code, handlers of radiation-generating equipment
shall
will
meet the following radiation safety
requirements
obligations:
(1)
Software-based technique selections, a chart, or a combination of the two
shall
will be
provided in the vicinity of the radiation-generating equipment's control panel
which specifies, for examinations performed with that system, the following
information:
(a) Patient's body part,
radiographic projection, anatomical size or age, and the technique factors to
be utilized for each;
(b) Type and
size of the image receptor to be used;
(c) Type and focal distance of the grid to be
used, if any; and
(d)
Source-to-image receptor distance (SID) to be used, except for fluoroscopy, and
dental intraoral or panoral radiography;
(2) Gonadal shielding of not less
than 0.5 millimeter lead equivalent material shall be used for human patients,
who have not passed the reproductive age, during radiologic procedures in which
the gonads are in the useful beam, except for cases in which this would
interfere with the radiologic procedure;
(3)(2)
Except for patients who cannot be moved out of the room, only the staff,
ancillary personnel or other persons
required
needed for
the medical procedure or training
shall
will be in the room during the radiologic procedure.
Other than the patient being examined:
(a) All
individuals shall
will be positioned such that no part of the body
shall
will be
struck by the useful beam unless protected by not less than 0.5 millimeter lead
equivalent material;
(b) The x-ray
operator, other staff, ancillary personnel, and other persons
required
needed for the medical procedure
shall
will be
protected from the direct scatter radiation by protective aprons or whole body
protective barriers of not less than 0.25 millimeter lead equivalent material;
and
(c) Human patients who cannot
be removed from the room shall
will be protected from the direct scatter radiation by
whole body protective barriers of not less than 0.25 millimeter lead equivalent
material or shall
will be so positioned that the nearest portion of the
body is at least two meters (6.5 feet) from both the tube head and the nearest
edge of the image receptor;
(4)(3) If performing a
radiologic procedure requires auxiliary support for holding a patient or an
image receptor, the handler
shall
will ensure the following:
(a) Mechanical holding devices
shall
will be
used when the procedure permits their use in lieu of having an individual hold
the patient or image receptor;
(b)
Written safe operating procedures
required
obligated by
paragraph (B)(4) of rule
3701:1-66-04 of the
Administrative Code
shall
will indicate the
requirements
obligations for selecting someone to hold a patient or
image receptor, and the procedure that
shall
will be
followed. All individuals holding a patient or image receptor during radiation
exposures
shall
will be at least eighteen years of age; and
(c) No individual shall
will routinely
hold patients or image receptors during radiologic procedures;
(5)(4) The facility
shall
will
have protective aprons and gloves available in sufficient numbers to provide
protection to anyone who is involved with x-ray operations;
(6)(5)
Any radiation worker participating in fluoroscopic, veterinary, or mobile or
portable x-ray procedures
shall be
required
is obligated to wear an
individual monitoring device unless
the registrant
demonstrates
the IRRP or radiation expert
determines it is unlikely the radiation worker will receive in excess of
the doses specified in paragraphs (B)(1)(a) to (B)(1)(c) of rule
3701:1-38-14 of the
Administrative Code;
(7)(6) The entrance air
kerma resulting from the technique used for the specified average adult human
patient for routine diagnostic radiography
shall
will not exceed
the values listed in table 2. The entrance air kerma resulting from the
technique used for routine intraoral bitewing exams
shall
will not exceed
the values listed in table 3. All values of entrance air kerma are specified as
free-in-air, without backscatter. The corresponding entrance exposure in
milliroentgens is listed in parentheses. Linear extrapolation or interpolation
shall
will be
used for an x-ray tube potential (kVp) not listed in table 3.;
|
Table 2.
|
|
Radiographic technique
|
Adult thickness cm
|
Entrance air kerma mGy (mR)
|
|
Chest (pa), (non-grid)
|
23
|
0.26 (30)
|
|
Chest (pa), (grid)
|
23
|
0.35 (40)
|
|
Abdomen (kub)
|
23
|
5.26 (600)
|
|
Lumbo-sacral spine (ap)
|
23
|
6.13 (700)
|
|
Thoracic spine (ap)
|
23
|
3.50 (400)
|
|
Full spine
|
23
|
3.50 (400)
|
|
Cervical spine (ap)
|
13
|
1.75 (200)
|
|
Skull (lateral)
|
15
|
1.75 (200)
|
|
Foot (dp)
|
8
|
0.88 (100)
|
|
Table 3.
|
|
Tube Voltage kVp
|
D-Speed Film mGy (mR)
|
F-Speed Film Digital Receptor mGy (mR)
|
|
50
|
4.82 (550)
|
2.45 (280)
|
|
55
|
4.56 (520)
|
2.19 (250)
|
|
60
|
4.12 (470)
|
1.93 (220)
|
|
65
|
3.64 (415)
|
1.66 (190)
|
|
70
|
3.15 (360)
|
1.45 (165)
|
|
75
|
2.72 (310)
|
1.23 (140)
|
|
80
|
2.28 (260)
|
1.01 (115)
|
|
85
|
2.06 (235)
|
0.92 (105)
|
|
90
|
1.84 (210)
|
0.83 (95)
|
|
95
|
1.71 (195)
|
0.74 (85)
|
|
100
|
1.58 (180)
|
0.61 (70)
|
(8)(7) Procedures and
auxiliary equipment designed to minimize patient and radiation worker exposure
shall
will be
utilized as follows:
(a) For facilities
utilizing radiographic film, the speed of the screen and film combinations used
shall
will be
the fastest speed consistent with the diagnostic objective of the examinations.
Film cassettes without intensifying screens shall
will not be used
for any routine diagnostic radiography, with the exception of veterinary and
specimen radiography;
(b)
Radiation-generating equipment subject to rule
3701:1-66-05 of the
Administrative Code
shall
will not be utilized in procedures where the
source-to-skin distance (SSD) is less than thirty centimeters, except for
veterinary x-ray systems;
(c) If
grids are used between the patient and the image receptor to decrease scatter
to the image receptor and improve contrast, the grid
shall
will be:
(i) Properly aligned, with the x-ray tube
side facing the correct direction, and the grid centered to the central ray;
and
(ii) The proper focal distance
for the SID being used;
(9)(8) Except for
radiation-generating equipment used for veterinary, portable, dental panoral,
dental intraoral, lithotripsy, or bone densitometry applications, the operator
shall stand
will
be behind a protective barrier, either in a separate room, in a protected
booth, or behind a shield and be able to see the
patient without leaving the protected barrier; and
(10)(9)
Each radiographic image, or a record linked with each radiographic image,
shall
will
contain the following:
(a) Patient
identification;
(b) Date of
examination; and
(c) Operator
identification.
(H) In addition to other applicable
structural shielding
requirements
obligations in Chapter 3701:1-66 of the Administrative
Code, handlers of radiation-generating equipment
shall
will:
(1) For all units, except those used for bone
densitometry, mammography, dental panoral or dental intraoral radiography:
(a) Use a radiation expert to prepare a
shielding design to include specifications for all structural radiation
barriers:
(i) Prior to new construction, or
renovation; and
(ii) For new
radiation-generating equipment installations which might cause a significant
increase in radiation hazard.
(b) Prior to patient use, use a radiation
expert to determine compliance with exposure levels in accordance with rule
3701:1-38-14 of the
Administrative Code by performing:
(i) An area
radiation survey for new installation of radiation-generating
equipment.
(ii) An area radiation
survey for reinstallation or after any change in structural shielding unless,
in the documented determination of a radiation expert, the reinstallation or
change will not cause a significant increase in radiation hazard.
(c) Use a radiation expert to
perform a re-calculation of area radiation survey results after any increase in
clinical workload that exceeds the assumptions used in the existing radiation
survey.
(d) Obtain a written report
of the shielding design and the area radiation survey. A copy of the report
shall
will be
made available to the department's inspector upon request.
(2) Assure that no individual operates or
permits the operation of radiation-generating equipment unless structural
shielding and protective barriers are used such that no person other than the
patient being examined
shall
will receive a total effective dose equivalent in
excess of the limits prescribed in rules
3701:1-38-12 and
3701:1-38-13 of the
Administrative Code.
(3) Provide a
protective barrier either in a separate room, in a protected booth, or use a
mobile barrier that will intercept the useful beam and any direct scattered
radiation.
(4) Provide a window of
lead equivalency affording protection equal to that required
obligated by
the adjacent barrier, a television monitoring system, or a mirror system large
enough and so placed that the operator can see the patient without having to
leave the protected area during exposure.
(5) Assure the stationary CT and mobile CT
radiation-generating equipment used in a fixed location provides for two-way
aural communications between the human patient and operator.
(I) Notwithstanding paragraph
(H)(1)(b)(ii) of this rule, reinstallation of radiation-generating equipment of
the same operating parameters, location and geometry does not
require
obligate another area radiation survey as long as the
previous documented area radiation survey is maintained and available for
inspection.
(J) In addition to all
applicable rules in Chapter 3701:1-66 of the Administrative Code, handlers of
radiation-generating equipment
shall
will meet the following quality assurance
requirements
obligations:
(1)
X-ray systems and associated components used on humans and certified pursuant
to 21 C.F.R. part
1020 (as effective on the effective date of this rule)
shall
will be
maintained in compliance with applicable requirements of that standard, any
modifications to the original components or systems
must
will comply with
that standard, and handlers
shall
will maintain documentation of compliance between
inspections;
(2) The handler
shall
will
maintain the following information for all radiation-generating equipment for
inspection by the department:
(a) User's
manuals;
(b) Records of surveys,
calibrations, maintenance, and modifications performed on the
radiation-generating equipment which shall
will be
maintained between inspections; and
(c) A copy of all correspondence with the
department regarding each piece of radiation-generating equipment;
(3) Unless otherwise specified in
another rule in this chapter, each installation using a piece of
radiation-generating equipment and using analog image receptors, such as
radiographic film,
shall
will have available suitable equipment for handling
and processing radiographic images in accordance with the following provisions:
(a) For manually processing film:
(i) Developer and fixer tanks
shall
will be
constructed of mechanically rigid, corrosion resistant material; and
(ii) The temperature of solutions in the
tanks
shall
will be maintained within the range of 15.6 to 26.7
degrees Celsius (sixty to eighty degrees Fahrenheit). Film
shall
will be
developed in accordance with the time-temperature relationships recommended by
the film manufacturer, or in absence of such recommendations, with the
following time-temperature chart:
|
Time-Temperature Chart
|
|
Thermometer Reading
(Degrees)
|
Minimum Developing Time
(Minutes)
|
|
°C
|
°F
|
|
|
26.7
|
80
|
2
|
|
26.1
|
79
|
2
|
|
25.6
|
78
|
2.5
|
|
25.0
|
77
|
2.5
|
|
24.4
|
76
|
3
|
|
23.9
|
75
|
3
|
|
23.3
|
74
|
3.5
|
|
22.8
|
73
|
3.2
|
|
22.2
|
72
|
4
|
|
21.7
|
71
|
4
|
|
21.1
|
70
|
4.5
|
|
20.6
|
69
|
4.5
|
|
20.0
|
68
|
5
|
|
19.4
|
67
|
5.5
|
|
18.9
|
66
|
5.5
|
|
18.3
|
65
|
6
|
|
17.8
|
64
|
6.5
|
|
17.2
|
63
|
7
|
|
16.7
|
62
|
8
|
|
16.1
|
61
|
8.5
|
|
15.6
|
60
|
9.5
|
(iii)
Devices shall
will be utilized which will indicate the actual
temperature of the developer and signal the passage of a preset time
appropriate to the developing time required
needed;
(b) For automatic processors and other closed
processing systems:
(i) Films
shall
will be
developed in accordance with the time-temperature relationships recommended by
the film manufacturer; in the absence of such recommendations, the film
shall
will be
developed using the following chart:
|
Developer Temperature
(Degrees)
|
Minimum Immersion
Timea/
|
|
°C
|
°F
|
Seconds
|
|
35.5
|
96
|
19
|
|
35
|
95
|
20
|
|
34.5
|
94
|
21
|
|
34
|
93
|
22
|
|
33.5
|
92
|
23
|
|
33
|
91
|
24
|
|
32
|
90
|
25
|
|
31.5
|
89
|
26
|
|
31
|
88
|
27
|
|
30.5
|
87
|
28
|
|
30
|
86
|
29
|
|
29.5
|
85
|
30
|
|
a/ Immersion
time only, no crossover time included.
|
(ii)
The specified developer temperature and immersion time
shall
will be
posted in the darkroom, on the automatic processor, or be readily available to
the operator;
(c)
Processing deviations from the requirements
obligations listed above shall
will be
documented by the handler in such manner that the requirements
obligations of this rule are shown to be met or
exceeded, such as with extended processing, and special rapid
chemistry;
(d) Film processing
solutions shall
will be prepared in accordance with the directions
given by the film manufacturer, and shall
will be
maintained in strength by replenishment or renewal so that full development is
accomplished within the time specified by the manufacturer; and
(4) Pass boxes, if provided,
shall
will be
so constructed as to exclude light from entering the darkroom when cassettes
are placed in or removed from the boxes, and shall
will incorporate
adequate shielding from stray radiation to prevent exposure of undeveloped
film;
(5) The darkroom
shall
will be
light tight and use proper safelighting such that any film which would produce
an optical density between one and two when exposed in a cassette to
x-radiation and then processed
shall
will:
(a) Not suffer
an increase in optical density greater than 0.1 when exposed in the darkroom
for two minutes with all safelights on; and
(b) Not suffer an increase in optical density
greater than 0.05 for mammography when exposed to the darkroom for two minutes
with all safelights on;
(6) Darkrooms typically used by more than one
individual shall
will provide a method to prevent accidental entry of
light while undeveloped films are being handled or processed;
(7) Film shall
will be stored
in a cool, dry place and shall
will be protected from exposure to stray radiation.
Film in open packages shall
will be stored
in a light-tight container. If used, daylight film handling boxes
shall
will
preclude fogging of the film;
(8)
Expired x-ray film shall
will not be used for diagnostic radiographs;
(9) Cassettes, intensifying screens, and
computed radiographic imaging plates
shall
will be:
(a) Cleaned according to manufacturer's
specifications or an alternate frequency approved and documented by a radiation
expert in the quality assurance program;
(b) Inspected for damage; and
(c) Replaced as necessary to assure
radiographs of good diagnostic quality;
(10) For those registrants employing computed
and digital radiography imaging systems, the following
shall
will
apply:
(a) If the computed radiography reader
is located in the same room as the radiation-generating equipment and it is not
behind a protective barrier, x-ray exposures shall
will not be made
during processing;
(b) Computed
radiography plates shall
will be processed as soon as possible after exposure,
not to exceed eight hours under any circumstances;
(c) Computed radiography plates
shall
will be
adequately shielded from stray radiation. Registrants
shall
will
develop a process that will ensure that computed radiography plates are used
frequently enough or erased at least weekly so as to produce diagnostic quality
images; and
(d) Facilities other
than dental, podiatric, and veterinary, and fluoroscopy, shall
will complete at least annually image quality
evaluations appropriate for the equipment as established by a radiation expert
or system manufacturer;
(11) Annual evaluation of the integrity of
all required
necessary protective apparel.
(K) Upon discovery of a medical event, the
handler
shall
will:
(1) Contact the
department regarding the medical event within one business day;
(2) Provide a written report, including the
analysis of the medical event, by a radiation expert to the department within
fifteen business days of the medical event. The written report
must
will
include:
(a) The handler or registrant's
name;
(b) The name of the
prescribing physician;
(c) A brief
description of the event including the body site, dose delivered and any
critical structures involved;
(d)
Why the event occurred;
(e) The
effect, if any, on the individual who received the medical event;
(f) Actions, if any, that have been taken, or
are planned, to prevent recurrence; and
(g) Certification that the handler notified
the individual, or the individual's responsible relative or guardian, and if
not, why not.
(3)
Provide a clinical summary to the prescribing physician and patient within
fifteen business days; and
(4)
Maintain record of the medical event as part of the patient's permanent medical
record.
(L) The written
report in paragraph (K)(2) of this rule shall
will not contain
the individual's name or any other information that could lead to the
identification of the individual.
Notes
Ohio Admin. Code
3701:1-66-02
Effective:
12/13/2024
Five Year Review (FYR) Dates:
9/25/2024 and
12/13/2029
Promulgated
Under: 119.03
Statutory
Authority: 3748.04
Rule
Amplifies: 3748.01,
3748.02,
3748.04,
3748.05,
3748.06,
3748.07,
3748.12,
3748.121,
3748.13,
3748.14,
3748.15,
3748.17,
3748.18,
3748.19,
3748.20,
3748.22,
3748.99
Prior
Effective Dates: 02/15/2001, 07/03/2006, 06/20/2010, 07/01/2014, 05/15/2015,
12/20/2019