As used in this rule, "radiographic
equipment" means radiation-generating equipment that is operated for medical
use. The requirements of this rule do not apply to
radiation-generating equipment used for dental intraoral or panoral,
mammography, bone densitometry, computed tomography, fluoroscopy or spot film
imaging, and equipment used in radiation therapy
simulators. In
addition to other applicable rules contained in Chapters 3701:1-38 and
3701:1-66 of the Administrative Code, handlers of radiographic equipment shall
comply with the following:
(A) General purpose radiographic equipment
shall meet the following equipment standards:
(1) A means shall be provided for limiting
the x-ray beam to the image receptor and area of clinical interest
.
;
(2) For radiographic equipment having a
variable x-ray field limitation device, the limitation device shall have means
for independent stepless adjustment of both the length and width of the x-ray
field
.
;
(3) For
radiographic equipment that employs a light field for visually defining the
perimeter of the x-ray field, the light source shall be functional and the
total misalignment of the edges of the visually defined field with the
respective edges of the x-ray field along either the length or width of the
visually defined field shall not exceed two per cent of the source-to-image
distance (SID)
.
;
(4) In
addition to the requirements of paragraphs (A)(1), (A)(2) and (A)(3) of this
rule stationary radiographic equipment having a variable x-ray field limitation
device shall have:
(a) A means provided to
indicate when the axis of the x-ray beam is perpendicular to the plane of the
image receptor;
(b) A means
provided to align the center of the x-ray field with respect to the center of
the image receptor to within two per cent of the SID, when the x-ray beam is
perpendicular to the plane of the image receptor;
(c) A means to indicate the SID to within two
per cent. If it is a fixed SID, the distance shall be indicated with a
permanent marking;
(d) The beam
limiting device indicate numerically the field size in the plane of the image
receptor to which it is adjusted and be accurate within two per cent of the
SID; and
(e) Compliance
measurements made to discrete SID's and image receptor dimensions in common
clinical use, or at any other specific dimensions at which the beam-limiting
device or its associated diagnostic x-ray system is uniquely designed to
operate
.
;
(5)
General
purpose
Radiographic
radiographic equipment designed with only one image
receptor size at a fixed SID shall have:
(a) A
means to limit the field at the plane of the image receptor to dimensions no
greater than those of the image receptor, and to align the center of the x-ray
field with the center of the image receptor within two per cent of the SID;
or
(b) A means to align the field
such that the x-ray field at the plane of the image receptor shall not extend
beyond any edge of the image receptor
.
;
(6)
General
purpose
Radiographic
radiographic equipment designed with multiple
removable or selectable fixed
aperatures
apertures
shall have:
(a) A means to limit the x-ray
field in the plane of the receptor so that such field does not exceed each
dimension of the image receptor by more than two per cent of the SID when the
axis of the x-ray beam is perpendicular to the plane of the image
receptor;
(b) A means to align the
center of the x-ray field with the center of the image receptor to within two
per cent of the SID or for equipment uniquely designed where the beam axis is
intended to be offset from the center of the image receptor, the x-ray field at
the plane of the image receptor shall not extend beyond the image receptor;
and
(c) Paragraphs (A)(6)(a) and
(A)(6)(b) of this rule may be met with either:
(i) An assortment of removable,
fixed-aperture, beam limiting devices sufficient to meet the requirement for
each combination of image receptor size and SID for which the unit is designed
with each such device having clear and permanent markings to indicate the image
receptor size and SID for which it is designed; or
(ii) A beam-limiting device having multiple
fixed apertures sufficient to meet the requirement for each combination of
image receptor size and SID for which the unit is designed. Permanent, clearly
legible markings shall indicate the image receptor size and SID for which each
aperture is designed and shall indicate which aperture is in position for
use
.
;
(7) If a positive beam limitation
, (PBL) device is used, it shall meet the following
additional requirements:
(a) The PBL shall
prevent the production of x-rays when any of the following conditions are met:
(i) Either the length or width of the x-ray
field in the plane of the image receptor differs from the corresponding image
receptor dimensions by more than three per cent of the SID, except as permitted
by paragraph (A)(7)(c) of this rule;
(ii) The sum of the length and width
differences as stated in paragraph (A)(7)(a)(i) of this rule without regard to
sign exceeds four per cent of the SID; or
(iii) The beam-limiting device is at an SID
for which PBL is not designed for sizing;
(b) Compliance with paragraph (A)(7)(a) of
this rule shall be determined:
(i) When the
equipment indicates that the beam axis is perpendicular to the plane of the
image receptor; and
(ii) No sooner
than five seconds after insertion of the image receptor;
(c) The PBL system shall be capable of
operation, at the discretion of the operator, such that the size of the field
may be made smaller than the size of the image receptor through stepless
adjustment of the field size. The minimum field size at a SID of one hundred
centimeters shall be equal to or less than five centimeters by five
centimeters;
(d) The PBL system
shall be designed such that if a change in image receptor does not cause an
automatic return to PBL function as described in paragraph (A)(7)(a) of this
rule, then any change of image receptor size or SID must cause an automatic
return; and
(e) The PBL system
shall function as described in paragraph (A)(7) of this rule whenever all the
following conditions are met:
(i) The image
receptor is inserted into a permanently mounted cassette holder;
(ii) The image receptor length and width are
less than fifty centimeters;
(iii)
The x-ray beam axis is within plus or minus three degrees of vertical in any
direction and the SID is ninety to one hundred thirty centimeters inclusive; or
the x-ray beam axis is within plus or minus three degrees of horizontal and the
SID is ninety to two hundred five centimeters inclusive;
(iv) The x-ray beam axis is perpendicular to
the plane of the image receptor to within plus or minus three degrees;
and
(v) Neither tomographic nor
stereoscopic radiography is being performed
.
;
(8) A device shall be
provided to terminate the exposure at a preset time interval, preset product of
current and time, preset number of pulses, or preset radiation exposure to the
image receptor
.
;
(9) For
radiographic equipment that provides manual exposure control, the operator
shall be able to terminate the exposure at any time unless:
(a) The exposure is 0.5 second or less;
or
(b) During serial radiography,
means are provided to permit completion of any single exposure of the series in
progress
.
;
(10) In the case of radiographic equipment
that provides automatic exposure control:
(a)
The control panel shall indicate when this mode of operation is
selected;
(b) The density setting
and automatic exposure control detector positions that are selected prior to
the exposure shall be indicated; and
(c) A visible signal shall indicate when an
exposure has been terminated at the back-up limit. Manual resetting shall be
required before further automatic timed exposures can be made
.
;
(11) The x-ray control panel shall
provide visual indication when x-rays are produced and an audible signal shall
indicate when the exposure has terminated
.
;
(12) The exposure control switch shall meet
the following requirements:
(a) The switch
shall be a "dead-man switch
;"
;
(b)
It shall not be possible to initiate an exposure when the timer is set to the
"zero" or "off" position if either position is provided; and
(c) The switch shall be permanently mounted
in a protected area so that it cannot be operated outside the protected area
except for portable, mobile, or veterinary
radiographic
radiation-generating equipment.
(B) In addition to the applicable
quality assurance requirements of Chapter 3701:1-66 of the Administrative Code,
handlers of
general purpose radiographic
equiment
equipment
shall comply with the following:
(1)
The kilovoltage peak (kVp) accuracy shall be within plus or minus ten per cent
of the indicated value
.
;
(2) The
accuracy of the timing device shall be within plus or minus ten per cent of the
indicated setting. The timing device shall be tested at a minimum of two
settings within the operative range of fifty milliseconds to one thousand
milliseconds
.
;
(3) The
coefficient of variation:
(a) Of the kVp
reproducibility for at least four consecutive exposures shall not exceed
0.05;
(b) Of the timing device
reproducibility for at least four consecutive exposures shall not exceed 0.05;
and
(c) Of radiation exposure
reproducibility for at least four consecutive exposures shall not exceed 0.05
for any specific combination of selected technique factors
.
;
(4) For radiographic equipment
having independent selection of x-ray tube current (mA), the average ratios of
exposure to the indicated mA-seconds (mAs) product obtained at any two
consecutive tube current settings shall not differ by more than ten per cent of
their sum
.
;
(5) For radiographic equipment
having a combined x-ray tube current-exposure time product, or mAs selector,
but not a separate tube current, or mA selector, the average ratios of exposure
to the indicated milliampere-seconds product (milligray/mAs) values obtained at
any two consecutive mAs selector settings shall not differ by more than ten per
cent of their sum
.
;
(6) The
average exposure ratio for paragraphs (B)(4) and (B)(5) of this rule shall be
expressed as follows:
|X1 -
X2| < 0.10 (X1+X2)
Where the value of X1 and X2 are the average milligray/mAs
values obtained at each of the two consecutive tube mA or mAs settings, or at
two settings differing by no more than a factor of two where the mA or mAs
selector provides continuous selection.
(C) In addition to the applicable radiation
safety rules in Chapter 3701:1-38 and rules
3701:1-66-02
and
3701:1-66-04
of the Administrative Code, the operator of
general
purpose radiographic equipment shall limit the useful beam to the area of
clinical interest, not to exceed the size of the image receptor by more than
two per cent of the source-to-image distance.
(D) Handlers of mobile or portable
radiographic equipment
radiation-generating equipment shall not be required
to comply with the requirements of paragraph
(I)
(H) of rule
3701:1-66-02
of the Administrative Code, and shall comply with the following:
(1) Mobile and portable
systems
radiation-generating equipment which are:
(a) Used continuously for greater than one
week in the same location, such as a room or suite, shall have the x-ray
control permanently mounted in a protected area so that the operator is
required to remain in that protected area during the entire exposure;
or
(b) Used for less than one week
at the same location shall be provided with either a protective barrier at
least 6.5 feet high for operator protection during exposures, or means shall be
provided to allow the operator to be at least six feet from the tube housing
assembly during exposures and the operator shall wear a protective apron of not
less than 0.25 millimeter lead equivalent when making exposures;
(2) Radiation emitted from the
x-ray tube when a capacitor energy storage system is fully charged and the
exposure switch, timer, or any discharge mechanism is not activated shall not
exceed an air kerma of 0.26 microgray in one minute at five centimeters from
any accessible surface of the diagnostic source assembly, with the
beam-limiting device fully open; and
(3) A tube stand or other mechanical support
shall be used so that the x-ray tube housing assembly shall not be hand-held
during exposures.
(E)
Handlers of stationary veterinary
radiographic
radiation-generating equipment shall not be required
to comply with the requirements of paragraph
(I)(4)
(H)(4) of rule
3701:1-66-02
of the Administrative Code. However, stationary veterinary
radiographic
radiation-generating equipment shall be provided with
either a 6.5 foot high protective barrier for operator protection during
exposures, or shall be provided with means to allow the operator to be at least
six feet from the tube housing assembly during exposures. If the operator or
assistant is not behind the protective barrier, a lead apron of not less than
0.25 millimeter lead equivalent shall be worn when making exposures
;
.
(F) Handlers of mobile or portable veterinary
radiographic
radiation-generating equipment shall not be required
to comply with the requirements of paragraph
(I)
(H) of rule
3701:1-66-02
of the Administrative Code. However, mobile or portable veterinary
radiographic
radiation-generating equipment shall be provided with
either a 6.5 foot high protective barrier for operator protection during
exposures, or shall be provided with means to allow the operator to be at least
six feet from the tube housing assembly during exposures. If the operator or
assistant is not behind the protective barrier, a lead apron of not less than
0.25 millimeter lead equivalent shall be worn when making exposures
;
.
(G) Handlers of certified veterinary
radiation-generating equipment specifically designed by the manufacturer to be
hand-held during radiographic exposures shall not be required to comply with
the requirements of paragraph
(I)
(H) of rule
3701:1-66-02
of the Administrative Code. The handler shall develop and implement safe
operating procedures as part of the quality assurance program specified in rule
3701:1-66-04
of the Administrative Code, which shall address and document at least the
following:
(1) Hand-held radiation-generating
equipment shall be used for intraoral, extremity or small animal purposes
only;
(2) Examination specific
source-to-image distances shall be developed and implemented to assure the
useful beam is limited to the area of clinical interest or no larger than the
image receptor;
(3) Operators of
the hand-held radiation-generating equipment and individuals participating in
the x-ray procedure shall be protected from direct scatter radiation by
protective aprons of not less than 0.25 millimeter lead equivalent
material;
(4) If the hand-held
radiation-generating equipment is designed with a back scatter shield, the
backscatter shield shall be in place during all radiographic
exposures;
(5) Storage and security
procedures shall be developed and implemented to assure hand-held
radiation-generating equipment is secured against unauthorized use or removal
when not under the control and constant surveillance of the handler;
(6) Hand-held radiation generating equipment
shall not be used in hallways or waiting rooms; and
(7) Operator training shall include
documented specific instruction to the x-ray operator regarding:
(a) The prohibition on placing any part of
their body into the useful beam unless protected by not less than 0.5
millimeter lead equivalent material;
(b) The proper use of source-to-image
distance for the examination to assure the size of the radiation beam is no
larger than necessary;
(c) Ensure
that all individuals required for the examination are wearing the appropriate
lead equivalent aprons and no bystanders are in the vicinity;
(d) The areas of use, proper storage and
security procedures for the hand-held radiation-generating equipment;
and
(e) The use of the dead-man
switch and software safety devices such as locks and sensors.
Notes
Ohio Admin. Code 3701:1-66-05
Effective:
12/20/2019
Five Year Review (FYR) Dates:
9/10/2019 and
05/01/2023
Promulgated
Under: 119.03
Statutory
Authority: ORC 3748.04
Rule
Amplifies: ORC 3748.01,
3748.02,
3748.04,
3748.05,
3748.06,
3748.07,
3748.12,
3748.121,
3748.13,
3748.14,
3748.15,
3748.17,
3748.18,
3748.19,
3748.20,
3748.22,
3748.99
Prior
Effective Dates: 02/15/2001, 02/10/2006, 04/01/2012,
07/01/2014