(1) Condition:
General Quality Control - The laboratory must establish and follow written
quality control procedures, including but not limited to electronic quality
control and internal quality control, for monitoring and evaluating the quality
of the analytical testing process of each method. General quality control shall
assure the accuracy and reliability of patient test results and reports. The
medical laboratory director shall review all general quality control or may
designate, in writing, the review to personnel meeting the qualification of
those respective positions. The medical laboratory director shall retain the
ultimate responsibility for the general quality control of the operation of the
laboratory. In addition, the laboratory must meet the applicable standards in
paragraphs
1200-06-03-.09(2)
through
1200-06-03-.09(11)
of this rule.
(2) Standard:
Laboratory Testing: For each test performed, the laboratory will be in
compliance if it
(a) Meets all applicable
quality control requirements specified in this rule; or
(b) Follows manufacturer's instructions when
using their products (instruments, kits, or test systems). In addition, the
laboratory must comply with requirements within any paragraph of this rule that
are unique to the laboratory facility and cannot be met by manufacturer's
instructions.
(3)
Standard: Test Methods, Equipment, Instrumentation, Reagents, Materials, and
Supplies. The laboratory must utilize test methods, equipment instrumentation,
reagents, materials, and supplies that provide accurate and reliable test
results and test reports.
(a) Test
methodologies and equipment must be selected and testing performed in a manner
that provides test results within the laboratory's stated performance
specifications for each test method as determined under Rule
1200-06-03-.09(4).
1. The laboratory must have appropriate and
sufficient equipment, instruments, reagents, materials, and supplies for the
type and volume of testing performed and for the maintenance of quality during
all phases of testing.
2. The
accuracy of analytical balance weights must be verified annually against
appropriate standard sources.
3.
The accuracy of thermometers must be verified annually against appropriate
standard sources or following the manufacturer's instructions.
4. The accuracy of pipettes must be verified
annually against appropriate standard sources or following the manufacturer's
instructions.
5. The accuracy of
centrifuge(s) must be verified annually against appropriate standard sources or
following manufacturer's instructions.
6. The accuracy of timers must be verified
annually against, appropriate standard sources or following manufacturer's
instructions.
(b) The
laboratory must define criteria for those conditions that are essential for
proper storage of reagents and specimens, for accurate and reliable test system
operation, and for test result reporting.
1.
These conditions include, if applicable, the following:
(i) Water quality;
(ii) Temperature;
(iii) Humidity; and
(iv) Protection of equipment and
instrumentation from fluctuations and interruptions in electrical current that
adversely affect patient test results and test reports.
2. Remedial actions taken to correct
conditions that fail to meet the criteria specified in this subparagraph must
be documented.
(c)
Reagents, solutions, culture media, control materials, calibration materials
and other supplies, as appropriate, must be labeled to indicate the following:
1. Identity and, when significant, titer,
strength or concentration;
2.
Recommend storage requirements;
3.
Preparation and expiration date;
4.
The date of receipt and date opened; and
5. Other pertinent information required for
proper use.
(d)
Reagents, solutions, culture media, control materials, calibration materials
and other supplies must be prepared, stored, and handled in a manner to ensure
the following:
1. Reagents, solutions, culture
media, controls, calibration materials and other supplies are not used when
they have exceeded their expiration date or when they have deteriorated or are
of substandard quality. The laboratory must comply with the FDA product dating
requirements of 21 CFR
610.53 for blood products and other
biologicals, and with the labeling requirement of
21 CFR
809.10 for all other in-vitro diagnostics.
Any exception to the product dating requirements in
21 CFR
610.53, will be granted by the FDA in the
form of an amendment of the product license, in accordance with
21 CFR
610.53(d). All exceptions
must be documented by the laboratory; and
2. Components of reagent kits of different
lot numbers are not interchanged unless otherwise specified by the
manufacturer.
(4) Standard: Procedure Manual.
(a) A written procedure manual for the
performance of all analytical methods used by the laboratory must be readily
available and followed by laboratory personnel. Textbooks may be used as
supplements to these written descriptions but may not be used in lieu of the
laboratory's written procedures for testing or examining specimens. Procedure
should be substantially in compliance with the CLSI, GP-2A, current version, or
any subsequent version.
(b) The
procedure manual must include, when applicable to the test procedure, all of
the following:
1. Requirements for specimen
collection and processing, and criteria for specimen rejection;
2. Procedures for microscopic examination,
including the detection of inadequately prepared slides;
3. Step-by-step performance of the procedure,
including test calculations and interpretation of results;
4. Preparation of slides, solutions,
calibrators, controls, reagents, stains and other materials using in
testing;
5. Calibration and
calibration verification procedures;
6. The reportable range for patient test
results as established or verified in Rule
1200-06-03-.09(5);
7. Control procedures;
8. Remedial action to be taken when
calibration or control results fail to meet the laboratory's criteria for
acceptability;
9. Limitations in
methodologies, including interfering substances;
10. Reference range (normal
values);
11. Imminent
life-threatening laboratory results or "critical values";
12. Current pertinent literature
references;
13. Appropriate
criteria for specimen storage and preservation to ensure specimen integrity
until testing is completed;
14. The
laboratory's system for reporting patient results including when appropriate,
the protocol for reporting critical values;
15. Description of the course of action to be
taken in the event that a test system becomes inoperable; and
16. Criteria for the referral of specimens
including procedures for specimen submission and handling as described in Rule
1200-06-03-.07(1)(a).
(c) Manufacturers' package inserts
or operator manuals may be used, when applicable, to meet the requirements of
parts (b)1. through (b) 13. of this paragraph. Information not provided by the
manufacturer must be provided by the laboratory.
(d) Procedures must be re-approved, signed
and dated by the medical laboratory director or designee.
(e) Procedures must be re-approved, signed
and dated if the directorship of the laboratory changes.
(f) Each change in a procedure must be
approved, signed, and dated by the current medical laboratory director or his
or her designee approved, in writing, by the medical laboratory
director.
(g) The laboratory must
maintain a copy of each procedure with the dates of initial use and
discontinuance. These records must be retained for two (2) years after a
procedure has been discontinued.
(5) Standard: Establishment and Verification
of Method Performance Specifications. Prior to reporting patient test results,
the laboratory must verify or establish, for each method, the performance
specifications for the following performance characteristics: accuracy;
precision; analytical sensitivity and specificity, if applicable; the
reportable range of patient test results; the reference range(s) (normal
values); and any other applicable performance characteristic.
(a) The provisions of this section are not
retroactive.
(b) A laboratory that
introduces a new procedure for patient testing must, prior to reporting patient
test results:
1. Verify or establish for each
method the performance specifications for the following performance
characteristics, as applicable:
(i)
Accuracy;
(ii) Precision;
(iii) Analytical sensitivity;
(iv) Analytical specificity to include
interfering substances;
(v)
Reportable range of patient test results;
(vi) Reference range(s); and
(vii) Any other performance characteristic
required for test performance.
(c) The laboratory must have
documentation of the verification or establishment of all applicable test
performance specifications.
(6) Standard: Equipment Maintenance and
Function Checks - The laboratory must perform equipment maintenance and
function checks to assure accurate and reliable test results and reports. This
includes electronic, mechanical and operational checks of equipment,
instruments, and test systems necessary for proper test performance and test
result reporting.
(a) Maintenance of
equipment, instruments, and test systems - For equipment, instruments or test
systems the laboratory must:
1. Establish a
maintenance protocol that ensures equipment, instrument, and test system
performance necessary for accurate and reliable test results and test result
reporting;
2. Perform maintenance
with at least the frequency specified in the above-mentioned maintenance
protocol; and
3. Document all
maintenance performed.
(b) Function checks of equipment,
instruments, and test systems - For equipment, instruments, or test systems the
laboratory must:
1. Define a function check
protocol that ensures equipment, instrument, and test system performance
necessary for accurate and reliable test results and test result
reporting;
2. Perform function
checks including background or baseline checks as specified in the
above-mentioned function check protocol. Function checks must be within the
laboratory's established limits before patient testing is conducted;
and
3. Document all function checks
performed.
(7) Standard: Calibration and Calibration
Verification Procedures - Calibration and calibration verification procedures
are required to substantiate the continued accuracy of the test method
throughout the laboratory's reportable range for patient test results.
Calibration is the process of testing and adjusting an instrument kit, or test
system to provide a known relationship between the measurement response and the
value of the substance that is being measured by the test procedure.
Calibration verification is the assaying of calibration materials in the same
manner as patient samples to confirm that the calibration of the instrument,
kit, or test system has remained stable throughout the laboratory's reportable
range for patient test results. The reportable range for patient test results
is the range of test result values over which the laboratory can establish or
verify the accuracy of instrument, kit or test system measurement response.
Calibration and calibration verification must be performed and documented as
required in this rule unless otherwise specified in paragraphs
1200-06-03-.09(12)
through
1200-06-03-.09(34).
For each method the laboratory must:
(a)
Perform calibration procedures
1. At a
minimum, in accordance with manufacturer's instructions, if provided, using
calibration materials provided or specified, as appropriate, and with at least
the frequency recommended by the manufacturer;
2. In accordance with criteria established by
the laboratory.
(i) Including the number,
type and concentration of calibration materials, acceptable limits for
calibration, and the frequency of calibration if manufacturer's instructions
are not provided; and
(ii) Using
calibration materials appropriate for the methodology and, if possible,
traceable to a reference method or reference material of known value;
and
3. Whenever
calibration verification fails to meet the laboratory's acceptable limits for
calibration verification; and
(b) Perform calibration verification
procedures
1. In accordance with the
manufacturer's calibration verification instructions when they meet or exceed
the requirements specified by the laboratory as described in Rule
1200-06-03-.09(7)(b)
2., or
2. In accordance with
criteria established by the laboratory
(i)
Including the number, type, and concentration of calibration materials,
acceptable limits for calibration verification and frequency of calibration
verification; and
(ii) Using
calibration materials appropriate for
(I) The
methodology and, if possible, traceable to a reference method or reference
material of known value; and
(II)
Verifying the laboratory's established reportable range of patient test
results, which must include at least a minimal (or zero[0]) value, a mid-point
value, and a maximum value at the upper limit of that range; and
(iii) At least once every six (6)
months and whenever any of the following occur:
(I) A complete change of reagents for a
procedure is introduced, unless the laboratory can demonstrate that changing
reagent lot numbers does not effect the range used to report patient test
results, and control values are not adversely affected by reagent lot number
changes.
Note: If reagents are obtained from a manufacturer and all of
the reagents for a test are packaged together, the laboratory is not required
to perform calibration verification for each package of reagents, provided the
packages of reagents are received in the same shipment and contain the same lot
number;
(II) There is major
preventive maintenance or replacement of critical parts that may influence test
performance;
(III) Controls reflect
an unusual trend or shift or are outside of the laboratory's acceptable limits
and other means of assessing and correcting unacceptable control values have
failed to identify and correct the problem; or
(IV) The laboratory's established schedule
for verifying the reportable range for patient test results requires more
frequent calibration verification than specified in Rule
1200-06-03-.09(7)(b)
2.(i), (ii), or (iii) and
(c) Document
all calibration verification procedures performed.
(8) Standard: Control Procedures - Control
procedures are performed on a routine basis to monitor the stability of the
method or test system; control materials provide a means to indirectly assess
the accuracy and precision of patient test results. Control procedures must be
performed as defined in this paragraph unless otherwise specified in paragraphs
1200-06-03-.09(12)
through
1200-06-03-.09(34).
(a) Control samples must be tested in the
same manner as patient specimens.
1. For
qualitative tests, the laboratory must include a positive and negative control
with each run of patient specimens.
2. For quantitative tests, the laboratory
must include at least two (2) samples of different concentrations of either
calibration materials, control materials, or a combination thereof with the
frequency determined in subparagraph
1200-06-03-.09(3)(a),
but not less frequently than once each run of patient specimens.
3. For electrophoretic determinations
(i) At least one (1) control sample must be
used in each electrophorectic cell; and
(ii) The control sample must contain
fractions representative of those routinely reported in patient
specimens.
4. Each day
of use, the laboratory must evaluate the detection phase of direct antigen
systems using an appropriate positive and negative control material (organism
or antigen extract). When direct antigen systems include an extraction phase,
the system must be checked each day of use using a positive organism.
5. If calibration materials and control
materials are not available, the laboratory must have an alternative mechanism
to assure the validity of patient test results.
(b) Control samples must be tested in the
same manner as patient specimens.
(c) When calibration or control materials are
used, statistical parameters (e.g. mean and standard deviation) for each lot
number of calibration material and each lot of control material must be
determined through repetitive testing.
1. The
stated values of an assayed control material may be used as the target values
provided the stated values correspond to the methodology and instrumentation
employed by the laboratory and are verified by the laboratory.
2. Statistical parameters for unassayed
materials must be established over time by the laboratory through concurrent
testing with calibration materials or control materials having previously
determined statistical parameters.
(d) Control results must meet the
laboratory's criteria for acceptability prior to reporting patient test
results.
(e) Reagent and supply
checks.
1. The laboratory must check each lot
or shipment of reagents, discs, stains, antisera and identification systems
(systems using two (2) or more substrates) when prepared or opened for positive
and negative reactivity, as well as graded reactivity if applicable.
2. Each day of use (unless otherwise
specified in this rule), the laboratory must test staining materials for
intended reactively to ensure predictable staining characteristics.
3. The laboratory must check fluorescent
stains for positive and negative reactivity each time of use (unless otherwise
specified in this rule).
4. The
laboratory must check each batch or shipment of media for sterility, if it is
intended to be sterile and if sterility is required for testing. Media must
also be checked for its ability to support growth, and as appropriate,
selectivity/inhibition and/or biochemical response. The laboratory may use
manufacturer's control checks of media provided the manufacturer's product
insert specifies that the manufacturer's quality control checks meet the
current standards of the Clinical and Laboratory Standards Institute (CLSI) for
media quality control. The laboratory must document that the physical
characteristics of the media are not compromised and report any deterioration
in the media to the manufacturer. The laboratory must follow the manufacturer's
specifications for using the media and be responsible for the test
results.
5. A batch of media
(solid, semi-solid, or liquid) consists of all tubes, plates, or containers of
the same medium prepared at the same time and in the same laboratory, or, if
received from an outside source of commercial supplier, consists of all of the
plates, tubes or containers of the same medium that have the same lot numbers
and are received in a single shipment.
(f) For each method that is developed
in-house, that is a modification of the manufacturer's test procedure, the
laboratory must evaluate instrument and reagent stability and operator variance
in determining the number, type, and frequency of testing calibration or
control materials and establish criteria for acceptability used to monitor test
performance during a run of patient specimen(s). A run is an interval within
which the accuracy and precision of a testing system is expected to be stable,
but cannot be greater than twenty-four (24) hours or less than the frequency
recommended by the manufacturer. For each procedure, the laboratory must
monitor test performance using calibration materials or a combination
thereof.
(9) Standard:
Remedial Actions - Remedial action policies and procedures must be established
by the laboratory and applied as necessary to maintain the laboratory's
operation for testing patient specimens in a manner that assures accurate and
reliable patient test results and reports. The laboratory must document all
remedial actions taken when any of the following occur:
(a) The laboratory shall take any remedial
action to correct a noted error in specimen collection or loss of
specimen.
(b) Test systems that do
not meet the laboratory's established performance specifications, as determined
in Rule
1200-06-03-.09(5)
include but are not limited to the following:
1. Equipment or methodologies that perform
outside of established operating parameters or performance
specifications;
2. Patient test
values that are outside of the laboratory's reportable range of patient test
results; and
3. The determination
that the laboratory's reference range for a test procedure is inappropriate for
the laboratory's patient population.
(c) Results of control and calibration
materials that fail to meet the laboratory's established criteria for
acceptability. All patient test results obtained in the unacceptable test run
or since the last acceptable test run must be evaluated to determine if patient
test results have been adversely affected. The laboratory must take the
remedial action necessary to ensure the reporting of accurate and reliable
patient test results;
(d) If the
laboratory cannot report patient test results within its established time
frames the laboratory must determine, based on the urgency of the patient
test(s) requested, the need to notify the appropriate individual of the delayed
testing; and
(e) Errors in the
reported patient test results are detected. The laboratory must
1. Promptly notify the authorized person
ordering or individual utilizing the test results of reporting
errors;
2. Issue corrected reports
promptly to the authorized person ordering the test or the individual utilizing
the test results; and
3. Maintain
exact duplicates of the original report as well as the corrected report for two
(2) years.
(f) There
must be a procedure of review in operation to verify highly unusual results
such as delta values and critical values. This review must be performed and
documented by the supervisor or other person designated in writing on a daily
basis to identify possibly erroneous tests.
(10) Standard: Quality Control Records - The
laboratory must document and maintain records of all quality control activities
specified in paragraphs
1200-06-03-.09(2)
through
1200-06-03-.09(34)
and retain records for at least two (2) years. Immunohematology quality control
records must be maintained for a period of no less than five (5) years. In
addition, quality control records for blood and blood products must be
maintained for a period not less than five (5) years. In addition, quality
control records for blood and blood products must be maintained for a period
not less than five (5) years after processing records have been completed, or
six (6) months after the latest expiration date, whichever is the later date,
in accordance with 21 CFR
6060.160(d).
(11) Condition: Quality Control - Specialties
and Subspecialties - The laboratory must establish and follow written policies
and procedures for an acceptable quality control program that include
verification and assessment of accuracy, measurement of precision and detection
of error for all analyses and procedures performed by the laboratory. In
addition to the general requirements specified in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11),
the laboratory must meet the applicable requirements of paragraphs
1200-06-03-.09(12)
through
1200-06-03-.09(34)
or an CMS approved equivalent procedure and the following:
(a) Meets quality control requirements
specified in this paragraph; or
(b)
Follows manufacturer's instructions when using products (instruments, kits, or
test systems) as well as specialty and subspecialty quality control.
(c) Failure to meet any of the applicable
conditions in paragraph
1200-06-03-.09(12)
through
1200-06-03-.09(34)
may result in revocation of licensure for the entire specialty or subspecialty
to which the condition applies, in accordance with Rule 12-6-3-.05.
(13) Condition:
Bacteriology - To meet the quality control requirements for bacteriology in
paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and in paragraphs
1200-06-03-.09(13)
through
1200-06-03-.09(17)
for the subspecialties for which it is certified under the specialty of
microbiology.
(a) The laboratory must check
positive and negative reactivity with control organisms as specified below:
1. Each day of use for catalase, coagulase,
beta-lactamase, and oxidase reagents and DNA probes;
2. Each week of use for Gram and acid-fast
stains, bacitracin, optochin, ONPG, X and V discs or strips; and
3. Each month of use for antisera.
(b) Each week of use, the
laboratory must check XV discs or strips with a positive control
organism.
(c) For antimicrobial
susceptibility tests, the laboratory must check each new batch of media and
each lot of antimicrobial discs before, or concurrent with, initial use, using
approved reference organisms.
1. The
laboratory's zone sizes or minimum inhibitory concentration for reference
organisms must be within established limits before reporting patient
results.
2. The laboratory must use
the appropriate control organism(s) to check the procedure as required by
current law under CLIA.
(14) Condition: Mycobacteriology - To meet
the quality control requirements for mycobacteriology, the laboratory must
comply with the applicable requirements in paragraph
1200-06-03-.09 (1) through
1200-63-.09(11) and with subparagraphs (a) through (d) of this paragraph. All
quality control activities must be documented.
(a) Each day of use, the laboratory must
check the iron uptake test with at least one (1) acid-fast organism that
produces a positive reaction and with an organism that produces a negative
reaction and check all other reagents or test procedures used for mycobacteria
identification with at least one (1) acid-fast organism that produces a
positive reaction.
(b) The
laboratory must check flurochrome acid-fast stains for positive and negative
reactivity each week of use.
(c)
The laboratory must check acid-fast stains each week of use with an acid-fast
organism that produces a positive reaction.
(d) For susceptibility tests performed on
Mycobacterium tuberculosis, the laboratory must comply with the applicable
requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(15) Condition: Mycology - To meet the
quality control requirements for mycology, the laboratory must comply with the
applicable requirements in paragraphs
1200-06-03-.09(1)
though
1200-06-03-.09(11)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(a) Each day
of use, the laboratory using the auxanographic medium for nitrate assimilation
must check the nitrate reagent with a peptone control.
(b) Each week of use, the laboratory must
check all reagents used with biochemical tests and other test procedures for
mycological identification with an organism that produces a positive
reaction.
(c) Each week of patient
testing, the laboratory must check acid-fast stains for positive and negative
reactivity.
(d) For susceptibility
tests, the laboratory must test each drug each day of use with at least one (1)
control strain that is susceptible to the drug. The laboratory must establish
control limits. Criteria for acceptable control results must be met prior to
reporting patient results.
(16) Condition: Parasitology - To meet the
quality control requirements for parasitology, the laboratory must comply with
the applicable requirements of paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (c) of this paragraph. All quality control
activities must be documented.
(a) The
laboratory must have available a reference collection of slides or photographs,
and, if available, gross specimens for identification of parasites and must use
these references in the laboratory for appropriate comparison with diagnostic
specimens.
(b) The laboratory must
calibrate and use the calibrated ocular micrometer for determining the size of
ova and parasites, if size is a critical parameter. Recalibration must be
performed if the ocular micrometer is moved to another microscope or objectives
are changed.
(c) Each month of use,
the laboratory much check permanent stains using a fecal sample control that
will demonstrate staining characteristics.
(17) Condition: Virology - To meet the
quality control requirements for virology, the laboratory must comply with the
applicable requirements in paragraph
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (c) of this paragraph. All quality control
activities must be documented.
(a) The
laboratory must have available host systems for the isolation of viruses and
test methods for the identification of viruses that cover the entire range of
viruses that are etiologically related to clinical diseases for which services
are offered.
(b) The laboratory
must maintain records that reflect the systems used and the reactions
observed.
(c) In tests for the
identification of viruses, the laboratory must simultaneously culture
uninoculated cells or cell substrate control as a negative control to detect
erroneous identification results.
(18) Condition: Diagnostic Immunology - The
laboratory must meet the applicable quality control requirements in paragraph
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and paragraphs 1200-3-.09(19) through
1200-06-03-.09(20)
for the subspecialties for which it is certified under the specialty of
diagnostic immunology.
(19)
Condition: Syphilis Serology - To meet the quality control requirements for
syphilis serology, the laboratory must comply with the applicable requirements
in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (c) of this paragraph. All quality control
activities must be documented.
(a) For
laboratories performing syphilis testing, the equipment, glassware, reagents,
controls, and techniques for tests for syphilis must conform to manufacturers'
specifications.
(b) The laboratory
must run serologic tests on patient specimens concurrently with a positive
serum control of known titer or controls of graded reactivity plus a negative
control.
(c) The laboratory must
employ positive and negative controls that evaluate all phases of the test
system to ensure reactivity and uniform dosages.
(d) The laboratory may not report test
results unless the predetermined reactivity pattern of the controls is
observed.
(e) All facilities
manufacturing blood and blood products for transfusion or serving as referral
laboratories for these facilities must meet the syphilis serology testing
requirements of 21 CFR
640.5(a).
(20) Condition: General Immunology
- To meet the quality control requirements for general immunology, the
laboratory must comply with the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(a) The
laboratory must run serologic tests on patient specimens concurrently with a
positive serum control of know titer or controls of graded reactivity and
uniform dosages when positive and negative controls are not
sufficient.
(b) The Laboratory must
employ controls that evaluate all phases of the test system (antigens,
complement, erythrocyte indicator systems, etc.) to ensure reactivity and
uniform dosages when positive and negative controls alone are not
sufficient.
(c) The laboratory may
not report test results unless the predetermined reactivity pattern of the
controls is observed.
(d) All
facilities manufacturing blood and blood products for transfusion or serving as
referral laboratories for these facilities must meet
1. The HIV testing requirements of 21 CFR
610.45; and
2. Hepatitis testing
and requirements of 21 CFR
610.40.
(22) Condition: Routine
Chemistry - To meet the quality control requirements for routine chemistry, the
laboratory must comply with the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11).
All quality control activities must be documented. In addition, for blood gas
analyses, the laboratory must:
(a) Calibrate
or verify calibration according to the manufacturer's instructions and with at
least the frequency recommended by the manufacturer;
(b) Test one (1) sample of control material
for each eight (8) hours of testing or as set out in the manufacturer's
instructions;
(c) Use a combination
of calibrators and control materials that include both low and high values on
each day of testing or as set out in the manufacturer's instruction;
and
(d) Include one (1) sample of
calibration material or control material each time patients are tested unless
automated instrumentation internally verifies calibration at least every thirty
(30) minutes or as set out in the manufacturer's instruction.
(23) Condition: Endocrinology - To
meet the quality control requirements for endocrinology, the laboratory must
comply with the applicable requirements contained in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11).
All quality control activities must be documented.
(24) Condition: Toxicology - To meet the
quality control requirements for toxicology, the laboratory must comply with
the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11).
However, this subsection shall not apply to drug testing for employment
purposes. All quality control activities must be documented. In addition, for
drug abuse screening using thin layer chromatography:
(a) Each plate must be spotted with at least
one (1) sample of calibration material containing all drug groups identified by
thin layer chromatography which the laboratory reports; and
(b) At least one (1) control sample must be
included in each chamber, and the control sample must be processed through each
step of patient testing, including extraction procedures.
(25) Condition: Hematology - To meet the
quality control requirements for hematology, the laboratory must comply with
the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(a) For
automated hematology testing systems, excluding coagulation, the laboratory
must include two (2) levels of controls for each twenty-four (24) hours of
operation.
(b) For all automated
coagulation testing systems, the laboratory must include two (2) levels of
control for each eight (8) hours of operation and each time a change in
reagents occurs.
(c) For manual
hemoglobin determinations (cyanmethemoglobin), the procedure must be
standardized with reference materials of known, certified values. At least four
(4) different hemoglobin concentrations must be used to prepare the calibration
curve or to calibrate the readout instruments. For procedures using calibration
curves, all curves must be repeated regularly and verified after serving or
recalibration of instruments. Photometer functions checks must be run and
recorded daily. Photometers must be checked for linearity periodically and when
instrument adjustments have been made with appropriate filters or
solutions.
(d) For manual
hematocrit determinations, the speed and timer of the microhematocrit
centrifuge must be checked at specific intervals. The constant packing time
(minimum spin to reach maximum packing of cells) must be assessed on
installation and reassessed when there has been a change in either the speed or
time.
(27) Condition: Cytology - To meet the
quality control requirements for cytology, the laboratory must comply with the
applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and subparagraphs (a) through (g) of this paragraph.
(a) The laboratory must assure that
1. All gynecologic smears are stained using a
Papanicolaou or modified Papanicolaou staining method;
2. Effective measures are taken to prevent
cross-contamination between gynecologic and nongynecologic specimens during the
staining process;
3. Nongynecologic
specimens that have a high potential for cross-examination are stained
separately from other nongynecologic specimens, and the stains are filtered or
changed following staining;
4.
Diagnostic interpretations are not reported on unsatisfactory smears;
and
5. All cytology slide
preparations are evaluated on the premises of a laboratory certified to conduct
testing in the subspecialty of cytology.
(b) The laboratory is responsible for
ensuring the following:
1. Each individual
engaged in the evaluation of cytology preparations by nonautomated microscopic
technique examines no more than one hundred (100) slides (one patient per
slide, gynecologic or nongynecologic, or both) in a twenty-four (24) hour
period, irrespective of the site or laboratory. This limit represents an
absolute maximum number of slides and is not being employed as a performance
target for each individual. Previously examined reactive , reparative,
atypical, premalignant or malignant gynecologic cases as defined in part (c)1.
of this paragraph, previously examined nongynecologic cytology preparations,
and tissue pathology slides examined by a pathologist are not included in the
one hundred (100) slide limit.
2.
For purposes of workload calculations, each slide preparation (gynecologic or
nongynecologic) made using automated, semi-automated, or other liquid-based
slide preparatory techniques which result in cell dispersion over one-half
(1/2) or less of the total available slide area and which is examined by
nonautomated microscopic technique counts as one-half (1/2) slide.
3. Records are maintained of the total number
of slides examined by each individual during each twenty-four (24) hour period,
irrespective of the site or laboratory, and the number of hours each individual
spends examining slides in the twenty-four (24) hour period; and
(i) The maximum number of one hundred (100)
slides described in part (b)1. of this paragraph may not be examined in less
than an eight (8) hour workday, and
(ii) For the purposes of establishing
workload limits for individual examining slides by nonautomated microscopic
technique on other than an eight (8) hour workday basis (includes full-time
employees with duties other than slide examination and part-time employees), a
period of eight (8) hours must be used to prorate the number of slides that may
be examined. The following formula shall be used to determine maximum slide
volume to be examined:
(iii)
(Number of hours examining slides x 100) - 8
(c) A pathologist must ensure the following:
1. All gynecologic smears interpreted to be
showing reactive or reparative changes, atypical squamous or glandular cells of
undetermined significance, or to be in the premalignant (dysplasia, cervical
intraepithelial neoplasia or all squamous intraepithelial lesions including
human papillomavirus-associated changes) or malignant category are confirmed by
a pathologist qualified by certification in anatomic pathology by the American
Board of Pathology. The report must be signed to reflect the review or, if a
computer report is generated with signature, it must reflect an electronic
signature authorized by the pathologist;
2. All nongynecologic cytologic preparation
are reviewed by the pathologist. The report must be signed to reflect the
pathologist's review or, if a computer report is generated with signature, it
must reflect an electronic signature authorized by the pathologist;
3. The slide examination performance of each
cytotechnologist is evaluated and documented, including performance evaluation
through the reexamination of normal and negative cases and feedback on the
reactive, reparative, a typical, malignant or premalignant cases as defined in
part (c)1. of this paragraph; and
4. A maximum number of slides, not to exceed
the maximum workload limit described in subparagraph (b) of this paragraph is
established by the pathologist for each individual examining slide preparations
by nonautomated microscopic technique.
(i)
The actual workload limit must be documented for each individual and
established in accordance with the individual's capability based on the
performance evaluation as described in part (c) 3. of this paragraph.
(ii) Records are available to document that
each individual's workload limit is reassessed at least every six (6) months
and adjusted when necessary.
(d) The laboratory must establish and follow
a program designed to detect errors in the performance of cytologic
examinations and the reporting of results.
1.
The laboratory must establish a program that includes a review of slides from
at least ten per cent (10%) of the gynecologic cases interpreted to be negative
for reactive, reparative, atypical premalignant or malignant conditions as
defined in part (c) 1. This review must be done by a pathologist qualified by
certification in anatomic pathology by the American Board of Pathology, a
cytology general supervisor qualified under Rule
1200-06-01-.23 or a
cytotechnologist qualified under Rule
1200-06-01-.24.
(i) The review must include negative cases
selected at random from the total caseload and from patients or groups of
patients that are identified as having a high probability of developing
cervical cancer, based on available patient information;
(ii) Records of initial examinations and
rescreening results must be available; and
(iii) The review must be completed before
reporting patient results on those cases selected.
2. The laboratory must compare clinical
information, when available, with cytology reports and must compare all
malignant and premalignant (as defined in part (c)1. of this paragraph)
gynecology reports with the histopathology report, if available in the
laboratory (either on-site or in storage), and determine the causes of any
discrepancies.
3. For each patient
with a current high grade intraepithelial lesion (moderate dysplasia or CIN-2
or above), the laboratory must review all normal or negative gynecologic
specimens received within the previous five (5) years, if available in the
laboratory (either on-site or in storage). If significant discrepancies are
found that would affect patient care, the laboratory must notify the patient's
physician and issue an amended report.
4. The laboratory must establish and document
an annual statistical evaluation of the number of cytology cases examined,
number of specimens processed by specimen type, volume of patient cases
reported by diagnosis (including the number reported as unsatisfactory for
diagnostic interpretation), number of gynecologic cases where cytology and
available histology are discrepant, the number of gynecologic cases where any
rescreen of a normal or negative specimen results in reclassification as
malignant or premalignant, as defined in part (c)1. of this paragraph, and the
number of gynecologic cases for which history results were unavailable to
compare with malignant or premalignant cytology cases as defined in part(c)1.
of this paragraph.
5. The
laboratory must evaluate the case reviews of each individual examining slides
against the laboratory's overall statistical values, document any
discrepancies, including reasons for the deviation, and document corrective
action, if appropriate.
(e) The laboratory report must:
1. Clearly distinguish specimens or smears,
or both, that are unsatisfactory for diagnostic interpretation; and
2. Contain narrative descriptive nomenclature
for all results.
(f)
Corrected reports issued by the laboratory must indicate the basis for
correction.
(g) The laboratory must
retain all slide preparations for five (5) years from the date of examination,
or slides may be loaned to proficiency testing programs, in lieu of maintaining
them for this time period, provided the laboratory receives written
acknowledgement of the receipt of slides by the proficiency testing program and
maintains the acknowledgement to document the loan of such slides.
Documentation for slides loaned or referred for purposes other than proficiency
testing must also be maintained. All slides must be retrievable upon
request.
(28) Condition:
Histopathology - To meet the quality control requirements for histopathology, a
laboratory must comply with the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and subparagraphs (a) through (c) of this paragraph. All quality control
activities must be documented.
(a) A control
slide of known reactivity must be included with each slide or group of slides
for differential, or special stains. Reaction(s) of the control slide with each
special stain must be documented each day of testing.
(b) The laboratory performing the
histopathology testing must retain stained slides at least ten (10) years from
the date of examination and retain specimen blocks at least five (5) years from
the date of examination.
(c) The
laboratory must retain remnants of tissue specimens in a manner that assures
proper preservation of the tissue specimens until the portions submitted for
microscopic examination have been examined and a diagnosis made by an
individual qualified by certification in anatomic pathology by the American
Board of Pathology.
(d) All tissue
pathology reports must be signed by an individual qualified as specified in
subparagraph (c) of this paragraph. If a computer report is generated with an
electronic signature, it must be authorized by the individual qualified as
specified in subparagraph (c) of this paragraph.
(e) The laboratory must utilize acceptable
terminology of a recognized system of disease nomenclature in reporting
results.
(29) Condition:
Oral Pathology - To meet the quality control requirements for oral pathology,
the laboratory must comply with the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and
1200-06-03-.09(28).
All quality control activities must be documented.
(30) Condition: Radiobioassay - To meet
quality control requirements for radiobioassay, the laboratory must comply with
the applicable requirements of paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11).
All quality control activities must be documented.
(31) Condition: Histocompatibility - In
addition to meeting the applicable requirements for general quality control in
paragraphs
1200-06-03-.09(1),
for quality control for general immunology in paragraph
1200-06-03-.09(20)
and for immunohematology in paragraph
1200-06-03-.09(33),
the laboratory must comply with the applicable requirements in subparagraphs
(a) through (d) of this paragraph. All quality control activities must be
documented.
(a) For renal allotransplantation,
the laboratory must meet the following requirements:
1. The laboratory must have available and
follow criteria for the following:
(i)
Selecting appropriate patient serum samples for crossmatching;
(ii) The technique used in
crossmatching;
(iii) Preparation of
donor lymphocytes for crossmatching; and
(iv) Reporting crossmatch results;
2. The laboratory must
(i) Have available results of final
crossmatches before an organ or tissue is transplanted, and
(ii) Make a reasonable attempt and document
efforts to have available serum specimens for all potential transplant
recipients at initial typing, for periodic screening, for preptransplantation
crossmatch and following sensitizing events, such as transfusion and transplant
loss;
3. The
laboratory's storage and maintenance of both recipient sera and reagents must
(i) Be at an acceptable temperature range for
sera and components;
(ii) Use a
temperature alarm system and have an emergency plan for alternate storage;
and
(iii) Ensure that all specimens
are properly identified and easily retrievable;
4. The laboratory's reagent typing sera
inventory (applicable only to locally constructed trays) must indicate source,
bleeding date and identification number, and volume remaining.
5. The laboratory must properly label and
store all labile components utilized in histocompatibility testing.
6. The laboratory must:
(i) HLA type all potential transplant
recipients;
(ii) Type cells from
organ donors referred to the laboratory; and
(iii) Have available and follow a policy that
establishes when antigen redefinition and retyping are required;
7. The laboratory must have
available and follow criteria for
(i) The
preparation of lymphocytes for HLA-A, B and DR typing;
(ii) Selecting typing reagents, whether
locally or commercially prepared;
(iii) The assignment of HLA antigens;
and
(iv) Assuring that reagents
used for typing recipients and donors are adequate to define all major and
International Workshop HLA-A, B and DR specificities for which reagents are
readily available;
8.
The laboratory must do the following:
(i)
Screen potential transplant recipient sera for performed HLA-A and B antibodies
with a suitable lymphocyte panel on sera collected
(I) At the time of the recipient's initial
HLA typing; and
(II) Thereafter,
following sensitizing events and upon request; and
(ii) Use a suitable cell panel for screening
patient sera (antibody screen), a screen that contains all the major HLA
specificities and common splits
(I) If the
laboratory does not use commercial panels, it must maintain a list of
individuals for fresh panel bleeding; and
(II) If the laboratory uses frozen panels, it
must have a suitable storage system,
9. The laboratory must check
(i) Each typing tray using
(I) Positive control sera;
(II) Negative control sera; and
(III) Positive controls for specific cell
types when applicable (i.e., T. cells, B cells, and monocytes); and
(ii) Each compatibility test (i.e.
mixed lymphocyte cultures, homozygous typing cells or DNA analysis) and typing
for disease-associated antigens using controls to monitor the test components
and each phase of the test system to ensure an acceptable performance
level;
10. Compatibility
testing for cellularly-defined antigens must utilize techniques such as the
mixed lymphocyte culture test, homozygous typing cells or DNA
analysis;
11. If the laboratory
reports the recipient's or donor's, or both, ABO blood group and D(Rho) typing,
the testing must be performed in accordance with
1200-06-03-.09(33);
12. If the laboratory utilizes immunologic
reagents (such as antibodies or complement) to remove contaminating cells
during the isolation of lymphocytes or lymphocyte subsets, the efficiency of
the methods must be verified with appropriate quality control
procedures;
13. At least once each
month, the laboratory must have each individual performing tests evaluate a
previously tested specimen as an unknown to verify his or her ability to
reproduce test results. Records of the results for each individual must be
maintained; and
14. The laboratory
must participate in at least one (1) national or regional cell exchange
program, if available, or develop an exchange system with another laboratory in
order to validate inter-laboratory reproducibility.
(b) If the laboratory performs
histocompatibility testing for:
1.
Transfusion and other non-renal transplantation, excluding bone marrow and
living transplant, all the requirements specified in this paragraph, as
applicable, except for the performance of mixed lymphocyte cultures, must be
met;
2. Bone marrow
transplantation, all the requirements specified in this paragraph, including
the performance of mixed lymphocyte cultures or other augmented testing to
evaluate class 11 compatibility, must be met; and
3. Non-renal solid organ transplantation, the
results of final crossmatches must be available before transplantation when the
recipient has demonstrated presensitization by prior serum screening except for
emergency situations. The laboratory must document the circumstances, if known,
under which emergency transplants are performed, and records must reflect any
information concerning the transplant provided to the laboratory by the
patient's physician.
(c)
Laboratories performing HLA typing for disease-associated studies must meet all
the requirements specified in this paragraph except for the performance of
mixed lymphocyte cultures, antibody screening and crossmatching.
(d) For laboratories performing organ donor
HIV testing the requirements of paragraph
1200-06-03-.09(20)
for the transfusion of blood and blood products must be met.
(32) Condition: Clinical
Cytogenetics - To meet the quality control requirements for clinical
cytogenetics, the laboratory must comply with the applicable requirement of
paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(11)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(a) When
determination of sex is performed by X and Y chromatin counts, these counts
must be based on an examination of an adequate number of cells. Confirmatory
testing such as full chromosome analysis must be performed for all atypical
results.
(b) The laboratory must
have records that reflect the media used and document the reactions observed,
number of cells counted, the number of cells karyotyped, the number of
chromosomes counted for each metaphase spread, and the quality of the banding
that the resolution is sufficient to support the reported results; and that an
adequate number of karyotypes are prepared for each patient.
(c) The laboratory also must have policies
and procedures for assuring an accurate and reliable patient sample
identification during the process of accessioning, cell preparation,
photographing or other image reproduction technique, and photographic printing,
and storage and reporting of results or photographs.
(d) The laboratory report must include the
summary and interpretation of the observations and number of cells counted and
analyzed and the use of recognized nomenclature, such as the International
System of Cytogenetic Nomenclature.
(e) Adequate slides, films, hard copies of
karyotypes and reports shall be retained for twenty (20) years.
(33) Condition: Immunohematology -
To meet the quality control requirements for immunohematology, the laboratory
must comply with the applicable requirements in paragraphs
1200-06-03-.09(1)
through
1200-06-03-.09(II)
and with subparagraphs (a) through (d) of this paragraph. All quality control
activities must be documented.
(a) The
laboratory must perform ABO group and D(Rho) typing, unexpected antibody
detection, antibody identification and compatibility testing in accordance with
manufacturer's instruction, if provided, and as applicable, with 21 CFR
606
(with the exception of 21 CFR
606.20a, Personnel) and 21 CFR
640, et.
seq.
(b) The laboratory must
perform ABO group by concurrently testing unknown red cells with anti-A and
anti-B grouping reagents. For confirmation of ABO group, the unknown serum must
be tested with known A, and B red cells.
(c) The laboratory must determine the D(Rho)
type by testing unknown red cells with anti-D(anti-Rho) blood grouping
reagent.
(d) If required in the
manufacturer's package insert for anti-D reagents, the laboratory must employ a
control system capable of detecting false positive D(Rho) test
results.
(34) Condition:
Transfusion services and bloodbanking - If a facility provides services for the
transfusion of blood and blood products, the facility must be under the
adequate control and technical supervision of a pathologist or other doctor of
medicine or osteopathy meeting the qualifications in Rule 1200-6-1-.20,
technical supervision in immunohematology. The facility must ensure that there
are facilities for procurement, safekeeping and transfusion of blood and blood
products and that blood and blood products must be available to meet the needs
of the physicians responsible for the diagnosis, management, and treatment of
patients. The facility meets this condition by complying with the standards in
paragraphs
1200-06-03-.09(34)(a) through
(f).
(a)
Standard: Immunohematological collection, processing, dating periods, labeling
and distribution of blood and blood products.
In addition to the requirements of parts 1. through 4. of
this subparagraph, the facility must also meet the applicable quality control
requirements in 1200-06-03-.09(1)
through 1200-06-03-.09(10).
1. Blood and blood product collection,
processing and distribution must comply with 21 CFR Part
640 and
21CFR Part
606, and the testing laboratory must meet the applicable requirements of 42 CFR
Part
493.
2. Dating of blood and
blood products must conform to
21 CFR
610.53.
3. Labeling of blood and blood products
conform to 21 CFR Part
606, Subpart G.
4. Policies to ensure positive identification
of a blood or blood product recipient must be established, documented and
followed.
(b) Standard:
Blood and blood product storage facilities
1.
The blood and blood products must be stored under appropriate conditions, which
include an adequate temperature alarm system that is checked and documented
daily.
(i) An audible alarm system, including
a chart recorder, must monitor proper blood and blood product storage
temperature continuously.
(ii)
Activation, to include high and low limits, of alarm system must be documented
periodically, no less frequently than quarterly.
(iii) A remote alarm system is required if
location where blood or blood products are stored is not continuously
staffed.
(iv) There must be a
written procedure for response to an alarm when temperature limits are
exceeded.
(v) There must be a
written procedure for response to an alarm indicating specific personnel to
respond and their appropriate action.
2. If blood is stored or maintained for
transfusion purposes outside of a monitored refrigerator, the facility must
ensure and document that storage conditions, including temperature, are
appropriate to prevent deterioration of the blood or blood product.
(c) Standard: Arrangement for
services - In the case of services provided outside the blood bank, the
facility must have an agreement reviewed and approved by the director that
governs the procurement, transfer, and availability of blood and blood
products.
(d) Standard: Provision
of testing - There must be provision for prompt ABO blood group, D(Rho) type,
unexpected antibody detection and compatibility testing in accordance with
1200-06-03-.09(34)
of this rule and the laboratory investigation of transfusion reactions, either
through the facility or under arrangement with an approved facility on a
continuous basis, under the supervision of a pathologist or other doctor of
medicine or osteopathy meeting the qualifications of
1200-06-01-.20.
(e) Standard: Retention of samples of
transfused blood and patient specimens.
1.
Samples of each unit of transfused blood and the patient specimens must be
retained for a minimum of seven (7) days for further testing in the event of
reactions according to the facility's established procedures.
2. The facility must promptly dispose of
blood not retained for further testing that has passed its expiration
date.
(f) Standard:
Investigation of transfusion reactions.
1. The
facility, according to its established procedures, must promptly investigate
all transfusion reactions occurring in all facilities for which it has
investigational responsibility and make recommendations to the medical staff
regarding improvements in transfusion procedures.
2. The facility must document that all
necessary remedial actions are taken to prevent future recurrence of
transfusion reactions and that all policies and procedures are reviewed to
assure that they are adequate to ensure the safety of individuals being
transfused within the facility.
3.
When a fatality occurs as a result of a complication of blood or blood
component transfusions, the Director, Office of Compliance and Biologics
Quality, Center for Biological Evaluation and Research (CBER), FDA, must be
notified within twenty-four (24) hours. Within seven (7) days after the
fatality, a written report of the investigation must be submitted to the
Director at 1401 Rockville Pike, Suite 200 N, Rockville, MD 20852-1448. A copy
of the report should be sent to the collecting facility if appropriate. See
AABB Accreditations Requirements Manual.